Aortic Aneurysm, Thoracic Clinical Trial
— VEVOOfficial title:
Valiant Evo US Clinical Trial
NCT number | NCT02652949 |
Other study ID # | 10219498DOC |
Secondary ID | |
Status | Completed |
Phase | N/A |
First received | |
Last updated | |
Start date | April 2016 |
Est. completion date | March 2, 2023 |
Verified date | March 2024 |
Source | Medtronic Cardiovascular |
Contact | n/a |
Is FDA regulated | No |
Health authority | |
Study type | Interventional |
The purpose of the Valiant Evo US Clinical Trial is to demonstrate the safety and effectiveness of the Valiant Evo Thoracic Stent Graft System in subjects with a descending thoracic aortic aneurysm (DTAA) who are candidates for endovascular repair.
Status | Completed |
Enrollment | 100 |
Est. completion date | March 2, 2023 |
Est. primary completion date | December 2017 |
Accepts healthy volunteers | No |
Gender | All |
Age group | 18 Years and older |
Eligibility | Inclusion Criteria: 1. Subject is =18 years old. 2. Subject understands and voluntarily has signed and dated the Informed Consent Form approved by the Sponsor and by the Ethics Committee for this study. 3. Subject presents a Descending Thoracic Aortic Aneurysm (DTAA) which is localized below the ostium of Left Subclavian Artery (LSA) and above the ostium of celiac trunk 4. Subject has a DTAA that is one of the following: 1. A fusiform aneurysm with a maximum diameter that: - is = 50 mm and/or: - is > 2 times the diameter of the non-aneurysmal thoracic aorta and/or: - is < 50 mm and has grown = 5 mm within previous 12 months 2. A saccular aneurysm or a penetrating atherosclerotic ulcer 5. Subject's anatomy must meet all of the following anatomical criteria as demonstrated on contrast-enhanced CT and/or on contrast-enhanced Magnetic Resonance Angiogram (MRA) obtained within four (4) months prior to implant procedure: 1. Proximal and distal non-aneurysmal aortic neck diameter measurements must be = 16 mm and = 42 mm; 2. Proximal non-aneurysmal aortic neck length must be = 20 mm (for FreeFlo configuration) and = 25 mm (for Closed Web configuration) distal to the left common carotid artery (LCCA). Note: Proximal aortic neck length may include covering the LSA (with or without discretionary revascularization) when necessary to optimize device fixation and maximize aortic neck length. If occlusion of the LSA ostium is required to obtain adequate neck length for fixation and sealing, transposition or bypass to the LSA may be warranted. 3. Distal non-aneurysmal aortic neck length must be = 20 mm 6. Subject has adequate arterial access site or can tolerate a conduit that allows endovascular access to the aneurysmal site with the delivery system of the appropriate sized device chosen for the treatment. Exclusion Criteria: 1. Subject has a life expectancy of less than 1 year 2. Subject is participating in another investigational drug or device study which would interfere with the endpoints and follow-ups of this study. 3. Subject is pregnant. 4. Subject requires planned placement of the covered proximal end of the stent graft to occur in zones 0 or 1. 5. Subject has a thoracic aneurysm with a contained rupture or localized at the anastomosis of a previous graft (pseudo-/false aneurysm). 6. Subject has a mycotic aneurysm. 7. Subject has a dissection (type A or B) or an intramural hematoma or an aortic rupture in addition to the thoracic aneurysm. 8. Subject requires emergent aneurysm treatment, e.g., trauma or rupture. 9. Subject has received a previous stent or stent graft or previous surgical repair in the ascending and/or descending thoracic aorta, and/or in the aortic arch. 10. Subject requires surgical or endovascular treatment of an infra-renal aneurysm at the time of implant 11. Subject has had previous surgical or endovascular treatment of an infra-renal aortic aneurysm. 12. Treatment with the Valiant Evo Thoracic Stent Graft would require intentional revascularization of the brachio-cephalic artery or the left common carotid artery or the celiac trunk. 13. Subject has had or plans to have a major surgical or interventional procedure within 30 days before or 30 days after the planned implantation of the Valiant Evo Thoracic Stent Graft. This does not include planned procedures that are needed for the safe and effective placement of the stent graft (i.e., carotid/subclavian transposition, carotid/subclavian bypass procedure). 14. Subject has a significant and/or circumferential aortic mural thrombus at either the proximal or distal attachment sites that could compromise fixation and seal of the implanted stent graft. 15. Subject has a connective tissue disease (e.g., Marfan's syndrome, aortic medial degeneration). 16. Subject has a bleeding diathesis or coagulopathy, or refuses blood transfusion. 17. Subject has had a Myocardial Infarction (MI) within 3 months of the procedure. 18. Subject has had a Cerebrovascular Accident (CVA) within 3 months of the procedure. 19. Subject has a known allergy or intolerance to the device materials 20. Subject has a known allergy to anesthetic drugs 21. Subject has a known hypersensitivity or contraindication to anticoagulants, or contrast media, which is not amenable to pretreatment. 22. Subject has active or systemic infection at the time of the index procedure. |
Country | Name | City | State |
---|---|---|---|
United States | Albany Medical Center | Albany | New York |
United States | University of Michigan Health System | Ann Arbor | Michigan |
United States | Emory University Hospital | Atlanta | Georgia |
United States | University of North Carolina | Chapel Hill | North Carolina |
United States | Carolinas Medical Center | Charlotte | North Carolina |
United States | University of Virginia Medical Center | Charlottesville | Virginia |
United States | Cleveland Clinic | Cleveland | Ohio |
United States | Ohio State University | Columbus | Ohio |
United States | OhioHealth Riverside Methodist Hospital | Columbus | Ohio |
United States | Duke University Medical Center | Durham | North Carolina |
United States | Inova Fairfax Hospital | Falls Church | Virginia |
United States | University of Florida | Gainesville | Florida |
United States | Memorial Hermann Texas Medical Center | Houston | Texas |
United States | University of Minnesota Medical Center | Minneapolis | Minnesota |
United States | Sentara Norfolk General Hospital | Norfolk | Virginia |
United States | University of Pennsylvania | Philadelphia | Pennsylvania |
United States | Abrazo Arizona Heart Institute | Phoenix | Arizona |
United States | The Heart Hospital Baylor Plano | Plano | Texas |
United States | Barnes Jewish Hospital | Saint Louis | Missouri |
United States | Stanford Hospital | Stanford | California |
United States | Medstar Washington Hospital Center | Washington | District of Columbia |
Lead Sponsor | Collaborator |
---|---|
Medtronic Cardiovascular |
United States,
Type | Measure | Description | Time frame | Safety issue |
---|---|---|---|---|
Primary | Composite Safety and Effectiveness Endpoint That is Based on the Percentage of Subjects Who Experienced (a) Access and/or Deployment Failures; and/or (b) Major Device Effect (MDE) Within 30 Days Post Index Procedure | MDEs include: device-related secondary procedures, device-related mortality, conversion to open surgery, thoracic aortic aneurysm rupture.
Access failure: Inability to insert device due to mechanical failure or anatomic exclusions of the femoral or iliac arteries. Deployment failure: Deployment failure due to subject anatomy or mechanical failure. Specifically, deployment of the stent graft from the delivery system. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the primary endpoint global cohort of 87 subjects. The poolability on the primary endpoint between US and OUS data were assessed using Fisher's exact test. |
30 Days | |
Secondary | Safety and Effectiveness Outcome | Safety and Effectiveness outcome measures between 0-30 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort of 100 subjects (53 US, 47 Outside US [OUS]).
Measures include: peri-operative mortality, adverse events (AE), major adverse events (MAE), serious adverse events (SAE), secondary procedures, loss of stent graft patency, and endoleaks. |
30 Days | |
Secondary | Safety and Effectiveness Outcome | Safety outcome measures between 0-183 days and Effectiveness outcome measures between 31-183 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Safety measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE) and secondary procedures. Effectiveness measures include loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month. |
6 month | |
Secondary | Safety and Effectiveness Outcome | Safety outcome measures between 0-365 days and Effectiveness outcome measures between 184-365 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Safety measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE) and secondary procedures. Effectiveness measures include loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month. |
12 Month | |
Secondary | Safety and Effectiveness Outcome | Safety outcome measures between 0-730 days and Effectiveness outcome measures between 366-730 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Safety measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE) and secondary procedures. Effectiveness measures include loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month. |
24 Month | |
Secondary | Safety and Effectiveness Outcome | Safety outcome measures between 0-1095 days and Effectiveness outcome measures between 731-1095 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Safety measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE) and secondary procedures. Effectiveness measures include loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month. |
36 Month | |
Secondary | Safety and Effectiveness Outcome | Safety outcome measures between 0-1460 days and Effectiveness outcome measures between 1096-1460 days post implant procedure. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Safety measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE) and secondary procedures. Effectiveness measures include loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month." |
48 Month | |
Secondary | Safety and Effectiveness Outcome | Safety and Effectiveness outcome measures between implant procedure and 60 months. Data from the Valiant Evo US trial (NCT02652949) and Valiant Evo International trial (NCT02625324) were combined to create the global cohort.
Measures include: all-cause mortality (ACM), aneurysm related mortality (ARM), major device effects (MDE), adverse events (AE), major adverse events (MAE), serious adverse events (SAE), secondary procedures, loss of stent graft patency, endoleaks, stent graft migration as compared to 1-month, and aneurysm expansion greater than 5 mm as compared to 1-month. |
60 Month |
Status | Clinical Trial | Phase | |
---|---|---|---|
Active, not recruiting |
NCT02777593 -
Evaluation of the GORE® TBE Device in the Treatment of Lesions of the Aortic Arch and Descending Thoracic Aorta, Zone 2
|
N/A | |
Completed |
NCT02876263 -
Noninvasive Neuromonitoring of Surgery of the Thoracic Aorta
|
N/A | |
Terminated |
NCT02538822 -
Risk of Rupture of Aneurysms of the Thoracic Ascending Aorta (ATA) From the Dynamic Imaging
|
N/A | |
Completed |
NCT04930172 -
TOtal tRansfemoral branCHed endovasCular tHoracoabdominal Aortic Repair Registry
|
||
Completed |
NCT03207568 -
RE-GENERATION: The Safety and Performance of the Relay Pro and Relay NBS Pro Stent-graft Devices in the European Union (EU)
|
N/A | |
Recruiting |
NCT03010514 -
A Registry Study on Genetics and Biomarkers of Thoracic Aortic Aneurysm/Dissection
|
N/A | |
Completed |
NCT02266342 -
GORE® TAG® Thoracic Endoprosthesis French Mandatory Registry
|
||
Completed |
NCT00757003 -
To Evaluate the Safety and Efficacy for GORE TAG Thoracic Endoprosthesis in the Treatment of Thoracic Aortic Disease
|
Phase 3 | |
Recruiting |
NCT03574311 -
Preoperative Intravenous Ferric Carboxymaltose and Placebo in the Treatment of Patients Undergoing Cardiac Surgery
|
Phase 4 | |
Suspended |
NCT03410420 -
Detection of Hypoxia in Human Thoracic Aorta Using Pimonidazole Hydrochloride
|
Phase 1 | |
Recruiting |
NCT05703893 -
Investigation of the Neurovegetative Pattern in Patients With Thoracic Aortic Aneurysms (TAA)
|
||
Enrolling by invitation |
NCT05800743 -
Evaluation of the GORE® Ascending Stent Graft
|
N/A | |
Completed |
NCT02253082 -
Vasculopathic Injury and Plasma as Endothelial Rescue - OCTAplas Trial (EudraCT no. 2014-000452-28)
|
Phase 4 | |
Recruiting |
NCT02323581 -
Endovascular Treatment of TAAA and Aortic Arch Aneurysms Using Fenestrated and Branched Stent Grafts
|
N/A | |
Active, not recruiting |
NCT03414866 -
Thoraflex Hybrid Post-Market Study
|
||
Recruiting |
NCT01985906 -
Safety and Efficacy of Multiple Overlapping Uncovered Stents for Pararenal Aortic Aneurysm Repair
|
N/A | |
Active, not recruiting |
NCT02818972 -
RelayPro Thoracic Stent-Graft in Subjects With Thoracic Aortic Aneurysms and Penetrating Atherosclerotic Ulcers
|
N/A | |
Completed |
NCT03479164 -
Development of Ultra-Low Dose CT Based Screening for Aortic Aneurysms
|
N/A | |
Completed |
NCT01775046 -
Valiant Thoracic Stent Graft With the Captivia Delivery System in the Treatment of Descending Thoracic Aortic Diseases (VALIANT CAPTIVIA France)
|
||
Recruiting |
NCT04814238 -
Minimally Invasive Aortic Root and Aorta surGery rEgistry
|