Clinical Trials Logo

Clinical Trial Details — Status: Recruiting

Administrative data

NCT number NCT04220177
Other study ID # SETALATECBACAN002
Secondary ID
Status Recruiting
Phase N/A
First received
Last updated
Start date December 11, 2019
Est. completion date March 31, 2022

Study information

Verified date January 2020
Source Latecba S.A.
Contact Ezequiel Klimovsky, MD
Phone 54 11 4952 1360
Email eklimovsky@quid-consulting.com
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

This is a Prospective, Open-label, Multicenter, and Non-randomized Clinical Study. The main purpose of this study is to establish the efficacy and safety of a medical device system, SETA LATECBA Stent Graft, intended for the treatment of EVAR of pararenal AAA (patients with complex anatomy, not eligible for other surgical procedure).Other objective is to evaluate the technical performance of the device SETA LATECBA Stent Graft.


Description:

This is a Prospective, Open-label, Multicenter, and Non-randomized Clinical Study to determine the Safety and Efficacy of the SETA LATECBA Stent Graft for endovascular repair therapy (EVAR) in Subjects with Abdominal Aortic Aneurysm (AAA)..The main purpose of this study is to establish the efficacy and safety of a medical device system, SETA LATECBA Stent Graft, intended for the treatment of Endovascular Aneurysm Repair (EVAR) of pararenal AAA (patients with complex anatomy, not eligible for other surgical procedure).

Secondary objective(s): The secondary objective is to evaluate the technical performance of the device SETA. LATECBA Stent Graft which includes: access to the implantation site, visualization, deployment of the stent- graft, sizing compatibility, withdrawal of the delivery system, and compatibility with ancillary equipment.


Recruitment information / eligibility

Status Recruiting
Enrollment 42
Est. completion date March 31, 2022
Est. primary completion date December 31, 2021
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

- Male and female adults older than 18 years old, willing to cooperate with the study. No other demographic restriction.

- Initial diagnosis of AAA according to the standard of care criteria following the guidance Clinical Practice Guidelines for Endovascular Abdominal Aortic Aneurysm Repair

- Medical history: Complete history of previous, present and concomitant conditions, and current treatments for other conditions Signed informed consent. If the patient cannot consent, the subject's legally acceptable representative should consent

Diagnosis of Abdominal Aortic Aneurysm (AAA) according to the following criteria:

1. Diagnosis of an abdominal aortic aneurysm with a diameter = 5.0 cm for males or = 4.5 cm for females.

2. Aneurysm diameter with a growth rate = 0.5 cm/ 6 months

3. The juxtarenal neck length between 1 mm and 10 mm

4. Neck diameter =16 to = 26 mm

5. Suprarenal aorta diameter = infrarenal aorta diameter

6. Proximal neck angulated = 60 degrees relative to the long axis of the aneurysm

7. Immediate suprarenal aorta angulated = 60 degrees relative to the immediate infrarenal neck.

8. Common iliac diameter 8 to 20 mm.

9. Common iliac length 25 mm

10. Common iliac angle 60 degrees

11. Not eligible for other standard EVAR surgical procedure

Exclusion Criteria:

- General Exclusion Criteria

1. Life expectancy less than 2 years

2. Pregnant or breastfeeding or planning on becoming pregnant within 60 months

3. Included in another investigative drug or device study/studies or planning to do it within the following 24 months

Medical Exclusion Criteria

1. Previous treatment of AAA

2. Known allergy to polyester, stainless steel, nitinol and or gold, teflon, nylon,

3. Known anaphylactic reaction to contrast media.

4. Any type coagulopathy untreated.

5. Genetic connective tissue diseases, such as Marfan or Ehlers-Danlos syndromes

6. Planned interventional or surgical procedure within 30 days before or after AAA repair.

7. Renal dysfunction: creatinine level over 1.7 mg/dl

8. Systemic infection or fever over 38°C

AAA Anatomy Exclusion Criteria

9. Significant occlusive disease, tortuosity, or calcification.

i) Significant thrombus in the fixation sites. j) Fungal aneurysm k) Leaking/ruptured or symptomatic aneurysm. l) Traumatic aneurysm. m) Concurrent aneurysm of the thoracic aorta. n) One or both renal arteries coming from the aneurysmal sac. o) Proximal neck diameter, measured outer wall to outer wall on a sectional image (CTA) > 26 mm in diameter or < 16 mm in diameter p) Proximal neck angle > 60 degrees relative to the long axis of the aneurysm q) Immediate suprarenal aorta angle > 60 degrees relative to the immediate infrarenal neck.

r) Immediate suprarenal aorta angle > 60 degrees relative to the immediate infrarenal neck s) Aortic diameter, measured inner wall to inner wall on a sectional image (CTA) < 15 mm at the bifurcation t) Iliac/femoral anatomy that is unsuitable for access. u) Iliac artery diameter, measured outer wall to outer Wall on a sectional image (CTA) > 20 mm or < 8 mm at distal fixation site.

v) Iliac artery distal fixation site < 10 mm in length. w) Indispensable inferior mesenteric artery (IMA)

Study Design


Intervention

Device:
SETA LATECBA Stent Grafts
ABK model is a bifurcated polyester graft partially corrugated which is sutured at the proximal end (renal) to a balloon expandable 316 L stainless steel stent. The proximal half part of the stent is uncovered by polyester graft providing the anchoring to the healthy aorta wall when the device is implanted, as well as the perfusion of the renal arteries. The remaining half part is covered by polyester graft providing sealing of the aneurysm when it is expanded upon on the graft fabric and aneurysm neck. The distal end of each branch has the anchorage site for sealing after the connection with the iliac extensions RIK. The diameter of the anchorage site varies with the different ABK codes. The end of one branch is cone-shaped after the anchoring diameter with the purpose of promoting insertion of iliac extension RIK. The connection to iliac arteries is achieved by completing the aortic trunk ABK with 2 iliac extensions RIK.

Locations

Country Name City State
Canada Centre Hospitalier Universitaire Sherbrooke Sherbrooke Montreal

Sponsors (1)

Lead Sponsor Collaborator
Latecba S.A.

Country where clinical trial is conducted

Canada, 

Outcome

Type Measure Description Time frame Safety issue
Primary EFFICACY: Number of participants with Endoleak Type I (measured by Angio CT Scan) Type I endoleaks is diagnosed when there is a gap between the stent graft and the vessel wall. Usually is a result of the failure of the stent graft to achieve a circumferential seal of the aneurism. It produces systemic pressurization of the aneurysm sac and increases the risk of sac rupture. 30 days
Primary SAFETY: mortality percentage of died subjects 30-day
Secondary SAFETY: Major Adverse Events Composite endpoint including Q-wave myocardial infarction, Congestive heart failure, Cardiac ischemia requiring intervention Renal failure requiring dialysis, Bowel obstruction, ischemia or fistula, Stroke, Paralysis, Aneurysm leak and Device related death 30-day
Secondary Incidence of post-operative complications Complications of aneurysm repair followed by secondary interventions 12 months
Secondary Clinical success Composite endpoint including Successful deployment at intended site, No migration, Aneurysm exclusion, no type III and/or IV endoleak, No graft infection, No graft thrombosis, No aneurysm rupture, No conversion to open repair, 30-day
Secondary Clinical success Composite endpoint including No migration, Aneurysm exclusion, no type I, III and/or IV endoleak, No graft thrombosis, No aneurysm rupture. No aneurysm diameter expansion >5 mm, No conversion to open repair 6 months
Secondary Clinical success Composite endpoint including No migration, Aneurysm exclusion, no type I, III and/or IV endoleak, No graft thrombosis, No aneurysm rupture. No aneurysm diameter expansion >5 mm, No conversion to open repair 12 months
Secondary Technical Success of stent-graft implantation and delivery system usage Composite endpoint including Assessment of system performance (Procedural times, Total blood loss, Deployment and device patency Ease of insertion, Visualization, Delivery system retrieval, Device integrity, Kinks, No type IV endoleak) Ancillary equipment needed, Adjunctive maneuvers (Balloon dilatation of aortic neck Balloon dilatation of iliac arteries, Additional stent or surgical procedure required, Eventual use of proximal cuff, Contrast media volume) 12 months
See also
  Status Clinical Trial Phase
Active, not recruiting NCT04227054 - Abdominal Aortic Aneurysm Sac Healing and Prevention of Endoleaks N/A
Active, not recruiting NCT03687489 - Safety and Efficacy Study of Abdominal Aortic Aneurysm Stent Graft System N/A
Active, not recruiting NCT03507413 - Metformin Therapy in Non-diabetic AAA Patients Phase 2/Phase 3
Not yet recruiting NCT05756283 - The PREHAAAB Trial: Multimodal Prehabilitation for Patients Awaiting Open Abdominal Aortic Aneurysm Repair N/A
Not yet recruiting NCT04089241 - Fusion of CT Angiography With 3D Contrast Ultrasound as a Method for Follow up for Endovascular Aneurysm Repair N/A
Active, not recruiting NCT02604303 - A Prospective Analysis on the Expansion Rates of Abdominal Aortic Aneurysms
Completed NCT02229006 - Sodium Fluoride Imaging of Abdominal Aortic Aneurysms N/A
Terminated NCT02469376 - Evaluation of a New Imagingtechnologie for Thrombosis Phase 1
Terminated NCT01843335 - Contrast-enhanced Ultrasound in Follow-up After Endovascular Aneurysm Repair N/A
Completed NCT02224794 - LIFE Study: Least Invasive Fast-Track EVAR
Terminated NCT01425242 - Study on Anti-inflammatory Effect of Anti-hypertensive Treatment in Patients With Small AAA's and Mild Hypertension N/A
Completed NCT01118520 - AARDVARK (Aortic Aneurysmal Regression of Dilation: Value of ACE-Inhibition on RisK) Phase 2
Completed NCT00746122 - Immediate Management of the Patient With Rupture : Open Versus Endovascular Repair N/A
Completed NCT03952780 - Korean Registry of Percutaneous EVAR With INCRAFT Stent Graft for the Treatment of Abdominalaortic Aneurysm (K-INCRAFT)
Completed NCT00583414 - Endovascular Exclusion of Abdominal Aortic Aneurysms in High Risk Patients N/A
Recruiting NCT05864560 - Ankura™ AAA, Cuff and AUI Stent Graft System Post-Market Clinical Follow-Up
Completed NCT01683084 - Study of the Effectiveness of Telmisartan in Slowing the Progression of Abdominal Aortic Aneurysms Phase 4
Active, not recruiting NCT03180996 - Global Fenestrated Anaconda Clinical sTudy
Completed NCT02493296 - The Effect of Surgery on Central Aortic Pressure & haEmodynamics Study