Respiratory Rate Clinical Trial
Official title:
ChroniSense Polso Wrist Worn Respiration Rate Validation Study
20 volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate of the Polso Monitoring system
20 volunteer test subjects will be entered into an accuracy study that is designed to
validate accuracy of the respiratory rate of the Polso Monitoring system.
A range of stable respiratory rates will be elicited from each volunteer test subject. The
rates will be approximately 5, 10, 15, 20, 25, 30, 35, 40 and 45 breaths per minute; with
some natural variation from these exact numbers.
Each subject will be instrumented with an open system mask that allows for measurement of the
end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor will be
used as the Accuracy Reference Device (Reference) in this study.
;
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT03832205 -
Validation of Respiratory Rate and Heart Rate Measurements by Capaciflectors Placed in Four Locations on the Chest
|
||
Completed |
NCT04506684 -
Reveal LINQ Respiration Clinical Study
|
||
Completed |
NCT03685227 -
A Closer Look at Yoga Nidra: Sleep Lab Analyses
|
N/A | |
Completed |
NCT04179279 -
Comparative Analysis of a Non-contact Respiratory and Heart Rate Monitor Vs. a Conventional Clinically Validated Reference Monitor
|
||
Completed |
NCT04077593 -
On the PhonE Respiratory rAte Study (OPERA)
|
||
Completed |
NCT05366049 -
The Effect of Methods Used in Oral Antipyretic Administration on the Vital Findings of the Child With Fever
|
N/A | |
Completed |
NCT03148873 -
RespiraSense Versus Capnography & Manual Counting
|
||
Recruiting |
NCT05263791 -
Evaluation the Non-Inferiority of Airmod to Capnostream™35
|
N/A | |
Completed |
NCT05246358 -
Respiratory Rate Validation Study - ChroniSense Polso
|
N/A | |
Recruiting |
NCT05812378 -
Respiratory Monitoring System Safety and Performance Study
|
||
Not yet recruiting |
NCT06292299 -
The PARS Study: Paediatric Advanced Respiratory Service Study - An Observational Diagnostic Feasibility Study
|
||
Recruiting |
NCT05022264 -
Remote Investigation and Assessment of Vital Signs
|
||
Completed |
NCT06086769 -
High Flow Oxygen Therapy Effect on Healthy Subjects
|
N/A | |
Completed |
NCT04104737 -
Spire Medical Health Tag Respiratory Rate Validation in Adults
|
N/A | |
Completed |
NCT04064645 -
Bora Band Respiration Rate Study
|
N/A | |
Completed |
NCT01472133 -
Validation of Respiration Rate Algorithms
|
N/A | |
Completed |
NCT03306537 -
Sensor Validation Study - Quality Assurance
|
||
Active, not recruiting |
NCT03750318 -
Validation Study on RENEW's Aingeal at KK Women's and Children's Hospital
|
N/A | |
Completed |
NCT03754231 -
Validation Study on RENEW's Aingeal at KK Women's and Children's Hospital
|
N/A | |
Completed |
NCT04346498 -
Can Kangarooing Small Babies on the Back of a Mother Keep Them Warm and Stable
|
N/A |