Radiation Exposure Clinical Trial
Official title:
A Prospective Study to Estimate Inadvertent Radiation Exposure From Strontium Isotopes in Patients Who Have Undergone CardioGen-82® PET MPI Scanning at Two Clinical Sites
This study is to be performed at two clinical sites where patients received CardioGen-82 during the time period January to June, 2011 and where patients were scanned chronologically near to the index patients (2 days before or after). Of these, it is planned that at least 30 patients will be entered in this trial. Index patient is the term applied to individuals who had radiation spectra indicative of Sr-82 and Sr-85 detected by hand-held survey meters.
This clinical study protocol will use a three tier approach consisting of the following
examinations:
- Patient examination with Survey Meter: The patient will first be tested at the clinical
sites using a survey meter. In those patients whose radiation count is ≥ 2 times above
background count using a standard survey meter, the presence of Sr-82 and/or Sr-85 will
be assessed through a quantitative reading from a calibrated portable isotope
identifier.
- Patient examination with Portable Isotope Identifier: This examination will be done by
use of a portable isotope identifier, which will be supplied by the Sponsor to the
study sites. An assessor, independent to the study site and will perform the
examination. In those patients whose radiation count is ≥ 2 times above background
count using a standard survey meter, the presence of Sr-82 and/or Sr-85 will be
assessed through a quantitative reading from a calibrated portable isotope identifier.
The reading will be forwarded along with the date of the reading, the date of the
patient's CardioGen-82® scan, and the total dose of CardioGen-82® administered to the
patient, to an independent medical expert, independent to the study site, who will
back-calculate an estimated dose of Sr-82 and/or Sr-85 from these data.
- Patient examination at Oak Ridge National Laboratory: Any patient whose estimated dose
of Sr-82 exceeds the specified limit of 0.02 microcuries per mCi Rb-82 administered,
and/or whose estimated dose of Sr-85 exceeds the specified limit of 0.2 microcuries per
mCi Rb-82 administered, will be asked to travel to Oak Ridge National Laboratory in
Tennessee, for accurate whole body scanning. Scan data from Oak Ridge National
Laboratories will be analyzed by one or more independent medical experts to accurately
back-calculate exposure to Sr-82 and/or Sr-85, and to estimate health hazard for the
patient.
;
Observational Model: Cohort, Time Perspective: Prospective
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT03649451 -
Reference Dose Levels During Fluoroscopically Guided Procedures in Operating Rooms
|
||
Completed |
NCT03305978 -
Pulmonary Nodule Detection: Comparison of an Ultra Low Dose vs Standard Scan.
|
N/A | |
Recruiting |
NCT04139096 -
Genetic Susceptibility to Radiation Induced Thyroid Cancer
|
||
Completed |
NCT01424774 -
Prospective Evaluation of Strontium in Patients After CardioGen-82 PET MPI Scanning
|
N/A | |
Completed |
NCT00378053 -
A Prospective Study of Radiation Exposure to Surgeons
|
N/A | |
Recruiting |
NCT05561439 -
Individual Dosimetric Monitoring of Workers During Interventional Radiology and Cardiology Procedures for Cardiologists and Radiologists in France
|
||
Completed |
NCT04078165 -
Suspended Personal Protection System Versus Conventional Protection (Aka Zero-Gravity vs Shield and Apron)
|
N/A | |
Recruiting |
NCT04694391 -
Genomic Study of Relapse Esophageal Cancer After Radiotherapy
|
||
Not yet recruiting |
NCT04403815 -
Radiation Exposure During Coronary Procedures According to Vascular Access
|
||
Recruiting |
NCT04285944 -
Operator Radiation Protection During Cardiac Catheterization Using Mavig X-ray Protective Drapes®.
|
N/A | |
Completed |
NCT04363190 -
Genomic Instability in Vascular Surgeons
|
||
Completed |
NCT04404257 -
Reduce Radiation Exposure in Fluoroscopic Interventions Evaluation
|
N/A | |
Recruiting |
NCT04678258 -
Zero Fluoroscopy Voltage Guided vs. Linear CTI Ablation
|
N/A | |
Recruiting |
NCT03386500 -
Safety Study of BMX-001 (Radio-protector) in Patients With Newly Diagnosed Anal Cancer
|
Phase 1/Phase 2 | |
Recruiting |
NCT05833516 -
Hyper Adduction of Right Radial Artery/Arm vs Left Drag Over Technique
|
||
Recruiting |
NCT03946280 -
Utility of 3D Navigation to Reduce Ionizing Radiation in Common Flutter Ablation. Study of Personnel and Patient Dosimetry
|
N/A | |
Completed |
NCT03985488 -
Radiation Exposure Assessment in Fluoroscopy
|
||
Completed |
NCT03605030 -
Reducing Radiation Exposure to Operators During Invasive Cardiac Procedures
|
N/A | |
Recruiting |
NCT02689908 -
Individualized Radiation Dose Control
|
N/A | |
Completed |
NCT05347290 -
Comparison Of Effective Radiation Shielding Between Rampart IC, M1128 Shield vs Conventional Apron And Shields In Clinical Interventional Procedures
|
N/A |