Clinical Trial Summary

A study to test the accuracy of SpO2 sensors with various patient monitors, co-oximeters, and modules.


Description:

The U-TruSignal with different sensors is feasible for noninvasive and continuous SpO2 monitoring and meets the specified measurement accuracy when compared to a co-oximetry gold reference.


Recruitment information / eligibility

Status Completed
Enrollment 63
Est. completion date February 15, 2013
Est. primary completion date February 15, 2013
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years to 55 Years
Eligibility Inclusion Criteria:

- Ability to provide written informed consent or have a legally authorized representative provide written informed consent

- Subjects who are 18-55 years of age

- Subjects must be willing and able to comply with study procedures

- HbCO<3% (for smokers only)

Exclusion Criteria:

- Subject or a legally authorized representative is unable to provide written inform consent.

- Subjects that are considered morbidly obese (defined as BMI > 39.5),

- Subjects with injuries, deformities or abnormalities that may prevent proper application of the device under test,

- Smokers with COHb levels >3% or subjects who have smoked within two (2) days. Subjects will be screened for COHb levels of >3% with a Masimo Radical 7 (Rainbow).

- Subjects with known respiratory conditions (uncontrolled asthma, head cold, flu, pneumonia / bronchitis, respiratory distress / respiratory or lung surgery, emphysema, COPD, lung disease),

- Subjects with known heart or cardiovascular conditions,

- Female subjects that are actively trying to get pregnant or are pregnant,

- Subjects taking blood thinners or medications with aspirin or other antiplatelet medication in them (applies only to arterial draw study),

- Subjects with Systolic blood pressure >140mmHg,

- Subjects with Diastolic blood pressure >100mmHg,

- Subjects with heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Pulse oximetry
Pulse oximetry measurement using SpO2 sensors, patient monitors, co-oximeters, and various modules.

Locations

Country Name City State
United States Clinimark Louisville Colorado

Sponsors (1)

Lead Sponsor Collaborator
GE Healthcare

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Accuracy Root Mean Square (ARMS) The Accuracy Root Mean Square (ARMS) calculation is used to measure accuracy of the test SpO2 device compared to reference SpO2 device. A scatterplot is created with the forehead sensor saturation on the y-axis and the measured blood saturation on the x-axis. The line of identity is drawn representing the ideal points, meaning that the forehead sensor saturation is always the same as the blood saturation. The dispersion of the actual data points around the line of identity can be measured using a statistical calculation called Arithmetic Root Mean Square (ARMS). The smaller ARMs the closer the data points lie around the line of identity, representing a more accurate sensor as accurate and appropriate. Data analysis follows ISO 80601-2-61, 2011, Annex EE and the FDA Guidance Document for Pulse Oximeters (FDA Draft Guidance, July 19, 2007). 60 minutes
See also
  Status Clinical Trial Phase
Active, not recruiting NCT02846974 - Calibration and Validation of High Quality Low-Cost 3D Printed Pulse Oximeter
Completed NCT03383757 - U-TruSignal SpO2 Testing in Neonates N/A
Recruiting NCT04233827 - Mobile Biosensor for Measuring Vital Signs in Healthcare and Home Settings
Completed NCT02987985 - Efficacy of Opioid-free Anesthesia in Reducing Postoperative Respiratory Depression in Children Undergoing Tonsillectomy Phase 3
Completed NCT05185427 - Pulse Oximetry and Skin Pigmentation - Peds
Terminated NCT03843489 - SpO2 Accuracy In Vivo Testing for Neonates & Infants N/A
Completed NCT01472133 - Validation of Respiration Rate Algorithms N/A
Enrolling by invitation NCT05681637 - SpO2 Difference Between Fingers of the Hand During Sequential Desaturation and Step Resaturation N/A
Completed NCT05657665 - Skin Tone Measurement Method Investigation for Pulse Oximetry
Completed NCT06454318 - Screening Moderate to Severe Obstructive Sleep Apnea With Wearable Device
Enrolling by invitation NCT05657288 - SpO2 Comparison Between Commercial Smartwatch and Medical-grade Pulse Oximeter During Hypoxia N/A
Not yet recruiting NCT05994417 - Beyond Race: Objectively Assessed Skin Color and Its Association With Pulse Oximeter Bias in Critically Ill Infants
Recruiting NCT05554510 - Pulse Oximetry Errors in Hospitalized Patients Across Varying Skin Pigmentation
Enrolling by invitation NCT06298422 - Oxygen Saturations Across Tones of Skin
Recruiting NCT06063148 - Neonatal Pulse Oximetry Disparities Due to Skin Pigmentation N/A