Pulse Oximetry Clinical Trial
Official title:
SpO2 System Accuracy Testing
Verified date | January 2019 |
Source | GE Healthcare |
Contact | n/a |
Is FDA regulated | No |
Health authority | |
Study type | Interventional |
A study to test the accuracy of SpO2 sensors with various patient monitors, co-oximeters, and modules.
Status | Completed |
Enrollment | 63 |
Est. completion date | February 15, 2013 |
Est. primary completion date | February 15, 2013 |
Accepts healthy volunteers | Accepts Healthy Volunteers |
Gender | All |
Age group | 18 Years to 55 Years |
Eligibility |
Inclusion Criteria: - Ability to provide written informed consent or have a legally authorized representative provide written informed consent - Subjects who are 18-55 years of age - Subjects must be willing and able to comply with study procedures - HbCO<3% (for smokers only) Exclusion Criteria: - Subject or a legally authorized representative is unable to provide written inform consent. - Subjects that are considered morbidly obese (defined as BMI > 39.5), - Subjects with injuries, deformities or abnormalities that may prevent proper application of the device under test, - Smokers with COHb levels >3% or subjects who have smoked within two (2) days. Subjects will be screened for COHb levels of >3% with a Masimo Radical 7 (Rainbow). - Subjects with known respiratory conditions (uncontrolled asthma, head cold, flu, pneumonia / bronchitis, respiratory distress / respiratory or lung surgery, emphysema, COPD, lung disease), - Subjects with known heart or cardiovascular conditions, - Female subjects that are actively trying to get pregnant or are pregnant, - Subjects taking blood thinners or medications with aspirin or other antiplatelet medication in them (applies only to arterial draw study), - Subjects with Systolic blood pressure >140mmHg, - Subjects with Diastolic blood pressure >100mmHg, - Subjects with heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia |
Country | Name | City | State |
---|---|---|---|
United States | Clinimark | Louisville | Colorado |
Lead Sponsor | Collaborator |
---|---|
GE Healthcare |
United States,
Type | Measure | Description | Time frame | Safety issue |
---|---|---|---|---|
Primary | Accuracy Root Mean Square (ARMS) | The Accuracy Root Mean Square (ARMS) calculation is used to measure accuracy of the test SpO2 device compared to reference SpO2 device. A scatterplot is created with the forehead sensor saturation on the y-axis and the measured blood saturation on the x-axis. The line of identity is drawn representing the ideal points, meaning that the forehead sensor saturation is always the same as the blood saturation. The dispersion of the actual data points around the line of identity can be measured using a statistical calculation called Arithmetic Root Mean Square (ARMS). The smaller ARMs the closer the data points lie around the line of identity, representing a more accurate sensor as accurate and appropriate. Data analysis follows ISO 80601-2-61, 2011, Annex EE and the FDA Guidance Document for Pulse Oximeters (FDA Draft Guidance, July 19, 2007). | 60 minutes |
Status | Clinical Trial | Phase | |
---|---|---|---|
Active, not recruiting |
NCT02846974 -
Calibration and Validation of High Quality Low-Cost 3D Printed Pulse Oximeter
|
||
Completed |
NCT03383757 -
U-TruSignal SpO2 Testing in Neonates
|
N/A | |
Recruiting |
NCT04233827 -
Mobile Biosensor for Measuring Vital Signs in Healthcare and Home Settings
|
||
Completed |
NCT02987985 -
Efficacy of Opioid-free Anesthesia in Reducing Postoperative Respiratory Depression in Children Undergoing Tonsillectomy
|
Phase 3 | |
Completed |
NCT05185427 -
Pulse Oximetry and Skin Pigmentation - Peds
|
||
Terminated |
NCT03843489 -
SpO2 Accuracy In Vivo Testing for Neonates & Infants
|
N/A | |
Completed |
NCT01472133 -
Validation of Respiration Rate Algorithms
|
N/A | |
Enrolling by invitation |
NCT05681637 -
SpO2 Difference Between Fingers of the Hand During Sequential Desaturation and Step Resaturation
|
N/A | |
Completed |
NCT05657665 -
Skin Tone Measurement Method Investigation for Pulse Oximetry
|
||
Completed |
NCT06454318 -
Screening Moderate to Severe Obstructive Sleep Apnea With Wearable Device
|
||
Enrolling by invitation |
NCT05657288 -
SpO2 Comparison Between Commercial Smartwatch and Medical-grade Pulse Oximeter During Hypoxia
|
N/A | |
Not yet recruiting |
NCT05994417 -
Beyond Race: Objectively Assessed Skin Color and Its Association With Pulse Oximeter Bias in Critically Ill Infants
|
||
Recruiting |
NCT05554510 -
Pulse Oximetry Errors in Hospitalized Patients Across Varying Skin Pigmentation
|
||
Enrolling by invitation |
NCT06298422 -
Oxygen Saturations Across Tones of Skin
|
||
Recruiting |
NCT06063148 -
Neonatal Pulse Oximetry Disparities Due to Skin Pigmentation
|
N/A |