Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT05359328
Other study ID # PR 2022-466
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date April 26, 2022
Est. completion date April 29, 2022

Study information

Verified date July 2022
Source Rhythm Diagnostic Systems
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of this clinical study is to validate the oxygen saturation (SpO2) accuracy of the RDS MultiSense® Pulse Oximetry during non-motion conditions over the range of 70-100% arterial oxygen saturation (SaO2) as compared to arterial blood samples assessed by measuring carbon monoxide (CO) bound to hemoglobin by CO-Oximetry. Additionally, data will be collected for heart rate accuracy as compared to reference ECG. The end goal is to provide supporting documentation for the SpO2 and heart rate accuracy validation for RDS MultiSense® Pulse Oximetry. A minimum of 10 healthy adult male and female participants, ranging in age and pigmentation from light to dark, will be enrolled in the study to meet the study design requirements defined by International Organization for Standardization ISO 80601-2-61:2017: corrected version 2018-02 and by the FDA's Guidance for Pulse Oximeters (March 4, 2013). The participants will have an arterial catheter placed in the radial artery to allow for simultaneous blood samples during stable plateaus of induced hypoxic levels. The investigational device will be placed on the placed on the thorax of the participants (patch on the upper back and external electrode on the right pectoral) per the instructions for use. Simultaneous data collection will be set up for the system under test and control oximeter.


Recruitment information / eligibility

Status Completed
Enrollment 10
Est. completion date April 29, 2022
Est. primary completion date April 29, 2022
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years to 50 Years
Eligibility Inclusion Criteria: - 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either - Participant must have the ability to understand and provide written informed consent - Participant is 18 to 50 years of age - Participant must be willing and able to comply with study procedures and duration - Participant is a non-smoker or who has not smoked within 2 days prior to the study Exclusion Criteria: - Participant is considered as being morbidly obese (defined as BMI >39.5) - Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites, tattoo, extensive moles, or freckles in the optical path which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow participants to participate if the condition is noted and would not affect the particular sites utilized.) - Female participants that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the participant is known to be not of child-bearing potential). - Smoker participants who have refrained will be screened for Carboxyhemoglobin (COHb) levels >3% as assessed with a reference devie (Masimo Radical 7) - Participants with known respiratory conditions (self-reported) - Participants with known heart or cardiovascular conditions (self-reported, except for blood pressure and ECG review) - Self-reported health conditions as identified in the Health Assessment Form (self-reported) - Participants with known clotting disorders (self-reported) - Participants with severe contact allergies to standard adhesives, silicon, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported) - Participants with severe allergies to iodine (only applicable if iodine is used) - Participants with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocaine) - Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio < 0.4) - Unwillingness or inability to remove colored nail polish from test digits. - Unwillingness to have chest or other test sites shaved - Other known health condition, should be considered upon disclosure in health assessment form

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Pulse Oximeter, Respiratory Rate, Heart Rate and Arterial line
Measurement of physiological parameters

Locations

Country Name City State
United States Clinimark LLC Louisville Colorado

Sponsors (1)

Lead Sponsor Collaborator
Rhythm Diagnostic Systems

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Validation of SpO2 accuracy performance of RDS MultiSense® Pulse Oximeter by comparison to CO-Oximeter The Accuracy Root Mean Square (Arms) value will be used as a metric. Reference CO-Oximetry will be used as the basis for comparison for the investigational device SpO2 readings. From beginning to end of controled hypoxia, up to 2 hours
Secondary Evaluation of heart rate accuracy performance of RDS MultiSense® by comparison to ECG-derived heart rate The reference for comparison of the investigational device heart rate will be heart rate values collected from an ECG monitor.Heart rate accuracy will be assessed by review of the readings less than ±5 Beats Per Minute or ±10% from the Reference heart rate values. From beginning to end of monitoring, up to 2 hours
See also
  Status Clinical Trial Phase
Recruiting NCT04498598 - Structural Modification In Supraglottic Airway Device N/A
Completed NCT05532670 - N600X Low Saturation Accuracy Validation
Enrolling by invitation NCT04106401 - Intravascular Volumes in Hypoxia During Antarctic Confinement N/A
Recruiting NCT05883137 - High-flow Nasal Oxygenation for Apnoeic Oxygenation During Intubation of the Critically Ill
Not yet recruiting NCT05817448 - Hypoxia-induced Autophagy in the Pathogenesis of MAP
Recruiting NCT02661152 - DAHANCA 30: A Randomized Non-inferiority Trial of Hypoxia-profile Guided Hypoxic Modification of Radiotherapy of HNSCC. Phase 3
Terminated NCT02801162 - Evaluation of Accuracy and Precision of a New Arterial Blood Gas Analysis System Blood in Comparison With the Reference Standard N/A
Not yet recruiting NCT02201875 - Intrinsic Periodic Pattern of Breathing N/A
Completed NCT02943863 - Regional Ventilation During High Flow Nasal Cannula and Conventional Nasal Cannula in Patients With Hypoxia N/A
Completed NCT01922401 - Inverse Ratio Ventilation on Bariatric Operation N/A
Completed NCT02105298 - Effect of Volume and Type of Fluid on Postoperative Incidence of Respiratory Complications and Outcome (CRC-Study) N/A
Active, not recruiting NCT01681238 - Goal-directed Therapy in High-risk Surgery N/A
Completed NCT01463527 - Using Capnography to Reduce Hypoxia During Pediatric Sedation N/A
Completed NCT01507623 - Value of Capnography During Nurse Administered Propofol Sedation (NAPS) N/A
Withdrawn NCT00638040 - The Gene Expression Studies of the Role of Tumor Microenvironments in Tumor Progression N/A
Active, not recruiting NCT06097754 - Intermittent Exogenous Ketosis (IEK) at High Altitude N/A
Completed NCT04589923 - The VISION-Acute Study
Completed NCT05044585 - Evaluation of RDS MultiSense® in Desaturation Analysis in Healthy Volunteers N/A
Completed NCT03659513 - The Effect of ECMO on the Pharmacokinetics of the Drugs and Their Clinical Efficacy
Completed NCT03221387 - Sleep and Daytime Use of Humidified Nasal High-flow Oxygen in COPD Outpatients N/A