Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT04873947
Other study ID # PR2020-376
Secondary ID
Status Completed
Phase
First received
Last updated
Start date August 25, 2020
Est. completion date August 28, 2020

Study information

Verified date April 2021
Source Belun Technology Company Limited
Contact n/a
Is FDA regulated No
Health authority
Study type Observational

Clinical Trial Summary

This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. Testing was conducted under normal office environment conditions.


Description:

Pulse oximetry monitoring is considered a standard physiological measurement and is used by clinicians in everyday situations to estimate arterial oxygen saturation. A pulse oximeter is a device that measures the oxygen saturation of arterial blood non-invasively. In general, pulse oximeters use two wavelength absorption spectrophotometry to measure oxygen saturation based on the amount of reflected or scattered radiation. The wavelengths are selected to provide the best separation of absorbencies of oxy-hemoglobin (O2Hb) and deoxy-hemoglobin (HHb) states. The ratio of the two absorbencies is used to calculate the oxygen saturation (SpO2) value. Because an arterial sample of blood is not required to make the measurement, the pulse oximeter can provide non-invasive real time information. The clinical use of pulse oximeters has reduced the frequency and necessity of invasive arterial blood sampling, and has helped to improve patient safety by providing information to clinicians about patients' oxygenation status. The purpose of this study was to evaluate the SpO2 accuracy performance of the Belun Ring Oximeter placed on the index fingers during non-motion conditions over the range of 70-100% SaO2 to arterial blood samples assessed by CO-Oximetry for an SpO2 validation. The primary goal, was to show the SpO2 accuracy performance for the investigational device below. As a secondary goal, the pulse rate accuracy performance for the Belun ring was calculated. It was expected that the Accuracy Root Mean Square (ARMS) performance would meet the required specification of ARMS 3.5% or lower in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification. For the pulse rate accuracy, the desired goal was to have an accuracy performance of an ARMS of 3. No risks or adverse device effects were expected. There were no contraindications for use in the proposed study / study population. The study was conducted in accordance to the code of federal regulations for non-significant medical device studies and applicable ISO 14155 (2nd edition 2011-02-01), applicable sections of ISO 80601-2-61, second edition 2017-12, Corrected version 2018-02, and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013).


Recruitment information / eligibility

Status Completed
Enrollment 10
Est. completion date August 28, 2020
Est. primary completion date August 28, 2020
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years to 50 Years
Eligibility Inclusion Criteria: - 10-15 Adults with a minimum of 3 males and a minimum of 3 females, with the balance made up of either - Subject must have the ability to understand and provide written informed consent - Subject is 18 to 50 years of age - Subject must be willing and able to comply with study procedures and duration - Subject is a non-smoker or who has not smoked within 2 days prior to the study Exclusion Criteria: - Subject is considered as being morbidly obese (defined as BMI >39.5) - Compromised circulation, injury, or physical malformation of fingers, wrist, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.) - Female subjects that are actively trying to get pregnant or are pregnant (confirmed by positive urine pregnancy test unless the subject is known to be not of child-bearing potential). - Smoker Subjects who have refrained will be screened for COHb levels >3% as assessed with a Masimo Radical 7 (Rainbow) - Subjects with known respiratory conditions such as: (self-reported) - uncontrolled / severe asthma, - flu, - pneumonia / bronchitis, - shortness of breath / respiratory distress, - unresolved respiratory or lung surgery with continued indications of health issues, - emphysema, COPD, lung disease - Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review) - hypertension: systolic >140mmHg, Diastolic >90mmHg on 3 consecutive readings (reviewed during health screen). - have had cardiovascular surgery - Chest pain (angina) - heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia (reviewed during health screen) - previous heart attack - blocked artery - unexplained shortness of breath - congestive heart failure (CHF) - history of stroke - transient ischemic attack - carotid artery disease - myocardial ischemia - myocardial infarction - cardiomyopathy - Self-reported health conditions as identified in the Health Assessment Form (self-reported) - diabetes, - uncontrolled thyroid disease, - kidney disease / chronic renal impairment, - history of seizures (except childhood febrile seizures), - epilepsy, - history of unexplained syncope, - recent history of frequent migraine headaches, - recent symptomatic head injury (within the last 2 months) - cancer / chemotherapy - Subjects with known clotting disorders (self-reported) - history of bleeding disorders or personal history of prolonged bleeding from injury - history of blood clots - hemophilia - current use of blood thinner: prescription or daily use of aspirin - Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported) - Subjects with severe allergies to iodine (only applicable if iodine is used) - Subjects with severe allergies to lidocaine (or similar pharmacological agents, e.g. Novocain) - Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test (Ratio < 0.4) - Unwillingness or inability to remove colored nail polish from test digits. - Other known health condition, should be considered upon disclosure in health assessment form

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Pulse Oximeter
The investigational device will be placed on the finger of the left or right hand per the instructions for use. Simultaneous data collection will be set up for the system under test and control oximeter.

Locations

Country Name City State
United States Clinimark, LLC Louisville Colorado

Sponsors (2)

Lead Sponsor Collaborator
Belun Technology Company Limited Clinimark, LLC

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Collect SpO2 data for accuracy statistical analysis by 10 subjects Subjects were given medical grade mixtures of oxygen and nitrogen to induce stable plateaus across the range of 100% to 70%. The goal was to have an equal distribution of data by decade. The stable plateaus allowed data collection in the following SaO2 ranges 95-100, 90-95, 85-90, 80-85, 75-80, 70-75. In general, 4 to 8 discrete points were collected at each of the levels. 3 days
Primary Accuracy data analysis Data analysis follows ISO80601-2-61:2017, Annex EE and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013). 3 days
See also
  Status Clinical Trial Phase
Recruiting NCT04498598 - Structural Modification In Supraglottic Airway Device N/A
Completed NCT05532670 - N600X Low Saturation Accuracy Validation
Enrolling by invitation NCT04106401 - Intravascular Volumes in Hypoxia During Antarctic Confinement N/A
Recruiting NCT05883137 - High-flow Nasal Oxygenation for Apnoeic Oxygenation During Intubation of the Critically Ill
Not yet recruiting NCT05817448 - Hypoxia-induced Autophagy in the Pathogenesis of MAP
Recruiting NCT02661152 - DAHANCA 30: A Randomized Non-inferiority Trial of Hypoxia-profile Guided Hypoxic Modification of Radiotherapy of HNSCC. Phase 3
Terminated NCT02801162 - Evaluation of Accuracy and Precision of a New Arterial Blood Gas Analysis System Blood in Comparison With the Reference Standard N/A
Completed NCT02943863 - Regional Ventilation During High Flow Nasal Cannula and Conventional Nasal Cannula in Patients With Hypoxia N/A
Not yet recruiting NCT02201875 - Intrinsic Periodic Pattern of Breathing N/A
Completed NCT02105298 - Effect of Volume and Type of Fluid on Postoperative Incidence of Respiratory Complications and Outcome (CRC-Study) N/A
Completed NCT01922401 - Inverse Ratio Ventilation on Bariatric Operation N/A
Active, not recruiting NCT01681238 - Goal-directed Therapy in High-risk Surgery N/A
Completed NCT01463527 - Using Capnography to Reduce Hypoxia During Pediatric Sedation N/A
Completed NCT01507623 - Value of Capnography During Nurse Administered Propofol Sedation (NAPS) N/A
Withdrawn NCT00638040 - The Gene Expression Studies of the Role of Tumor Microenvironments in Tumor Progression N/A
Active, not recruiting NCT06097754 - Intermittent Exogenous Ketosis (IEK) at High Altitude N/A
Completed NCT04589923 - The VISION-Acute Study
Completed NCT05044585 - Evaluation of RDS MultiSense® in Desaturation Analysis in Healthy Volunteers N/A
Completed NCT03659513 - The Effect of ECMO on the Pharmacokinetics of the Drugs and Their Clinical Efficacy
Completed NCT03221387 - Sleep and Daytime Use of Humidified Nasal High-flow Oxygen in COPD Outpatients N/A