Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT03439306
Other study ID # PR-2016-207
Secondary ID
Status Completed
Phase N/A
First received February 14, 2018
Last updated February 14, 2018
Start date March 15, 2017
Est. completion date March 17, 2017

Study information

Verified date February 2018
Source Belun Technology Company Limited
Contact n/a
Is FDA regulated No
Health authority
Study type Observational

Clinical Trial Summary

The purpose of this study is to evaluate the oxygen saturation accuracy performance of Belun Ring Oximeter placed on the index fingers during non-motion conditions


Description:

The purpose of this study is to evaluate the SpO2 accuracy performance of Belun Ring Oximeter placed on the index fingers during non-motion conditions over the range of 70-100% SaO2, arterial blood samples, assessed by CO-Oximetry. It is expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system will meet the required specification of ARMS of 3.5% or less allowed for reflectance technology in non-motion conditions for the range of 70 - 100% SaO2.

The Clinimark proprietary Desaturation Fixture with Automated Data Collection is a single limb blow by system used to deliver medical grade oxygen and nitrogen gas mixtures to induce various hypoxic levels in subjects at a slow steady rate allowing an automatic marking and collection of the Control or secondary Transfer reference Pulse Oximeter and other pulse oximetry systems at 1 second intervals.

The Control Pulse Oximeter, an FDA cleared device, is used to monitor the oxygen saturation levels real time throughout the study for subject safety and to target stable plateaus. This device is used to assess the stability of the data.

Multi-parameter monitor used during the study to observe a subject's vital signs including ECG tracing, heart rate, respiratory rate, end-tidal CO2 with capnograph, secondary monitor for the oxygen concentration being delivered to the subject.

A whole blood analyzer (CO-Oximeter) is used as the reference standard device for obtaining the functional SaO2 value from arterial blood samples obtained during the study.


Recruitment information / eligibility

Status Completed
Enrollment 11
Est. completion date March 17, 2017
Est. primary completion date March 17, 2017
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years to 50 Years
Eligibility Inclusion Criteria:

- Subject must have the ability to understand and provide written informed consent.

- Subject is 18 to 50 years of age.

- Subject must be willing and able to comply with study procedures and duration.

- Subject is a non-smoker or who has not smoked within 2 days prior to the study.

- Male or female of any race.

- Subject demographics include a range of skin pigmentations, including at least 2 darkly pigmented subjects or 15% of the subject pool, whichever is larger.

Exclusion Criteria:

- Subject is considered as being morbidly obese (defined as BMI large than 39.5).

- Compromised circulation, injury, or physical malformation of fingers, toes, hands, ears or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study. (Note: Certain malformations may still allow subjects to participate if the condition is noted and would not affect the particular sites utilized.)

- Females who are pregnant, who are trying to get pregnant, or have a urine test positive for pregnancy on the day of the study.

- Subjects with COHb levels larger than 3% as assessed with a Masimo Radical 7 (Rainbow).

- Subjects with known respiratory conditions such as (self-reported) uncontrolled / severe asthma, flu, pneumonia / bronchitis, shortness of breath / respiratory distress, respiratory or lung surgery, emphysema, COPD, lung disease.

- Subjects with known heart or cardiovascular conditions such as: (self-reported), except for blood pressure and ECG review) hypertension, have had cardiovascular surgery, Chest pain (angina), heart rhythms other than a normal sinus rhythm or with respiratory sinus arrhythmia, previous heart attack, blocked artery, unexplained shortness of breath, congestive heart failure (CHF), history of stroke, transient ischemic attack, carotid artery disease, myocardial ischemia, myocardial infarction, cardiomyopathy.

- Self-reported health conditions as identified in the Health Assessment Form (self-reported), diabetes, thyroid disease, kidney disease / chronic renal impairment, history of seizures (except childhood febrile seizures), epilepsy, history of unexplained syncope, recent history of frequent migraine headaches, recent head injury, cancer / chemotherapy.

- Subjects with known clotting disorders (self-reported) history of bleeding disorders or personal history of prolonged bleeding from injury, history of blood clots, hemophilia, current use of blood thinner: prescription or daily use of aspirin.

- Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported).

- Unwillingness or inability to remove colored nail polish from test digits.

- Other known health condition, should be considered upon disclosure in health assessment form.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Pulse Oximeter
The Pulse Oximeter are placed on the left and right Index fingers of each subject of the healthy adult volunteer group to evaluate the SpO2 accuracy performance.

Locations

Country Name City State
United States Clinimark, LLC Louisville Colorado

Sponsors (2)

Lead Sponsor Collaborator
Belun Technology Company Limited Clinimark, LLC

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Collect SpO2 data for accuracy statistical analysis by 11 subjects Subjects were given medical grade mixtures of oxygen and nitrogen to induce stable plateaus across the range of 100% to 70%. The stable plateaus allowed data collection in the following SaO2 ranges 95-100, 90-95, 85-90, 80-85, 75-80, 67-75. In general, 4 to 8 discrete points were collected at each of the levels. 3 days
Primary Accuracy data analysis Data analysis follows ISO80601-2-61:2011, Annex EE and Pulse Oximeters - Premarket Notifications Submissions [510(k)s] Guidance For Industry and Food and Drug Administration Staff (issued: March 4, 2013) 3 days
See also
  Status Clinical Trial Phase
Recruiting NCT04498598 - Structural Modification In Supraglottic Airway Device N/A
Completed NCT05532670 - N600X Low Saturation Accuracy Validation
Enrolling by invitation NCT04106401 - Intravascular Volumes in Hypoxia During Antarctic Confinement N/A
Recruiting NCT05883137 - High-flow Nasal Oxygenation for Apnoeic Oxygenation During Intubation of the Critically Ill
Not yet recruiting NCT05817448 - Hypoxia-induced Autophagy in the Pathogenesis of MAP
Recruiting NCT02661152 - DAHANCA 30: A Randomized Non-inferiority Trial of Hypoxia-profile Guided Hypoxic Modification of Radiotherapy of HNSCC. Phase 3
Terminated NCT02801162 - Evaluation of Accuracy and Precision of a New Arterial Blood Gas Analysis System Blood in Comparison With the Reference Standard N/A
Not yet recruiting NCT02201875 - Intrinsic Periodic Pattern of Breathing N/A
Completed NCT02943863 - Regional Ventilation During High Flow Nasal Cannula and Conventional Nasal Cannula in Patients With Hypoxia N/A
Completed NCT02105298 - Effect of Volume and Type of Fluid on Postoperative Incidence of Respiratory Complications and Outcome (CRC-Study) N/A
Completed NCT01922401 - Inverse Ratio Ventilation on Bariatric Operation N/A
Active, not recruiting NCT01681238 - Goal-directed Therapy in High-risk Surgery N/A
Completed NCT01463527 - Using Capnography to Reduce Hypoxia During Pediatric Sedation N/A
Completed NCT01507623 - Value of Capnography During Nurse Administered Propofol Sedation (NAPS) N/A
Withdrawn NCT00638040 - The Gene Expression Studies of the Role of Tumor Microenvironments in Tumor Progression N/A
Active, not recruiting NCT06097754 - Intermittent Exogenous Ketosis (IEK) at High Altitude N/A
Completed NCT04589923 - The VISION-Acute Study
Completed NCT05044585 - Evaluation of RDS MultiSense® in Desaturation Analysis in Healthy Volunteers N/A
Completed NCT03659513 - The Effect of ECMO on the Pharmacokinetics of the Drugs and Their Clinical Efficacy