Clinical Trial Summary

The purpose of this study it to evaluate the accuracy of heart rate, heart rate variability and breathing rate measurement with the use of Shen.AI Vitals software developed by MX Labs.


Description:

Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software). The measurements will be taken with various camera-equipped devices (smartphones, tablets, laptops) using either front camera (self-measurement) or rear camera (measurement made by someone else), in two different lighting conditions (artificial light or natural daylight), with various levels of device stability (fully stable or held in hand/s), and in two different body positions (sitting or supine). In total, each participant will have 8 measurements taken. Standard deviation of normal heartbeat intervals (SDNN) will be used as the main HRV index. Apart from the average HR (averaged over the whole 1-minute measurement), instantaneous values of HR (averaged over 10 s) will also be determined and displayed every 1 s by the Shen.AI VItals software. In all cases, reference measurements will be taken simultaneously (chest ECG for the calculation of average and instantaneous HR as well as HRV; impedance pneumography for the determination of BR). The reference measurements will be taken with Finapres Nova with the ECG+Resp module (Finapres Medical Systems B.V.). For each measured parameter, the results from both methods will be compared using the Bland-Altman analysis or non-parametric limits of agreement.


Recruitment information / eligibility

Status Completed
Enrollment 136
Est. completion date June 6, 2023
Est. primary completion date June 6, 2023
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: - age = 18 years - ability to read and write in Polish or English - ability to operate a smartphone, tablet and laptop without the need for glasses - ability to understand the study procedure - ability to communicate and follow the instructions of the study staff - written consent to participate in the study and to the processing of personal data by the research team and the sponsor of the study Exclusion Criteria: - failure to provide the required information in the participant's information survey - failure to comply with the study procedure or the instructions of the study stuff - extensive facial skin damage (including abrasions, wounds, burns) - a disease process involving a large part of the face that makes the video- based measurement impossible - extensive face dressing - extensive facial skin tattoo or face painting - marked deformation of the face (e.g. a tumor) - inability to keep the head stable in the required position during the measurement - persistent cough making it impossible to remain still during the measurement - respiratory dysfunction in the form of shortness of breath or irregular or shallow breathing - diagnosed arrhythmia (except sinus bradycardia or tachycardia) - moderate or severe anemia - diagnosed: heart failure, left ventricular systolic dysfunction, aortic stenosis or other diseases or disorders within the heart vascular system leading to decreased stroke volume, low pressure amplitude or the so-called paradoxical pulse

Study Design


Related Conditions & MeSH terms


Intervention

Diagnostic Test:
measurement of physiological parameters wih Shen.AI Vitals software on various camera-equipped devices (smartphones, tablets, laptops)
Resting heart rate (HR), heart rate variability (HRV) and breathing rate (BR) will be measured (estimated) with the use of remote photoplethysmography (1-minute face video analysis with Shen.AI Vitals software). Reference measurements will be taken simultaneously (chest ECG for the calculation of HR and HRV; impedance pneumography for BR).

Locations

Country Name City State
Poland Uniwersyteckie Centrum Wsparcia Badan Klinicznych we Wroclawiu Wroclaw Dolnoslaskie

Sponsors (1)

Lead Sponsor Collaborator
MX LABS spólka z ograniczona odpowiedzialnoscia oddzial w Polsce

Country where clinical trial is conducted

Poland, 

Outcome

Type Measure Description Time frame Safety issue
Primary Accuracy of heart rate (HR) measurement with the software under test compared with the ECG-based measurement In each participant, average HR will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Primary Accuracy of heart rate variability index (SDNN) measurement with the software under test compared with the ECG-based measurement In each participant, the standard deviation of normal heartbeat intervals (SDNN) will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Primary Accuracy of breathing rate (BR) measurement with the software under test compared with the impedance pneumography-based measurement In each participant, the breathing rate will be estimated from 1-minute video-based remote photoplethysmography of the facial skin (software under test) and calculated from simultaneously recorded impedance pneumography signal. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Primary Accuracy of instantaneous heart rate measurement with the software under test compared with the ECG-based measurement In each participant, multiple instantaneous heart rate values (averaged over 10 s) will be estimated from the software under test and calculated from simultaneously recorded ECG. Measurements will be made in resting conditions (seated position), with artificial lighting and a fully stabilized camera device.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held by someone else (compared with reference measurements). Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of HR using the rear camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held by someone else (compared with reference measurements). Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of SDNN using the rear camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held by someone else (compared with reference measurements). Each participant will have one video-based measurement performed with the software under test on a smartphone held by someone else (facial video measurement of BR using the rear camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate (HR) with the software under test on a smartphone held in hand/s (compared with reference measurements). Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of HR using the rear camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test on a smartphone held in hand/s (compared with reference measurements). Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of SDNN using the rear camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of breathing rate (BR) with the software under test on a smartphone held in hand/s (compared with reference measurements). Each participant will have one video measurement performed with the software under test on a smartphone held in hand/s (facial video measurement of BR using the rear camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate (HR) with the software under test in daylight conditions (compared with reference measurements). Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of HR using the front smartphone camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in daylight conditions (compared with reference measurements). Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of SDNN using the front smartphone camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of breathing rate (BR) with the software under test in daylight conditions (compared with reference measurements). Each participant will have one facial video measurement performed in daylight conditions (1-minute measurement of BR using the front smartphone camera).
For reference, BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate (HR) with the software under test in the supine position (compared with reference measurements). Each participant will have one facial video measurement performed in the supine position (1-minute measurement of HR using the front smartphone camera).
For reference, HR will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using non-parametric limits of agreement.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement of heart rate variability index (SDNN - the standard deviation of NN intervals) with the software under test in the supine position (compared with reference measurements). Each participant will have one facial video measurement performed in the supine position (1-minute measurement of SDNN using the front smartphone camera).
For reference, SDNN will be calculated from the simultaneously recorded ECG signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Accuracy of measurement breathing rate (BR) with the software under test in the supine position (compared with reference measurements). Each participant will have one facial video measurement performed in the supine position (1-minute measurement of BR using the front smartphone camera).
For reference BR will be obtained from the simultaneously recorded impedance pneumography signal.
The results from both methods will be compared using the Bland-Altman analysis.
day 1 (single visit, approximately 30 minutes)
Secondary Utility of the software under test for self-measurement of physiological parameters (assessed with a questionnaire). The utility of the software under test for self-measurement of physiological parameters (HR, SDNN, and BR) will be assessed based on of the average score in the participant's questionnaire completed by each participant after the test procedure (i.e. after all measurements in the given participant).
The scale will be 1-5, with higher scores meaning better outcome.
day 1 (after a single visit, approximately 30 minutes)
Secondary Utility of the software under test for the measurement of someone's physiological parameters (assessed with a questionnaire). The utility of the software under test for the video-based measurement of someone's physiological parameters will be assessed based on the average score in the participant's questionnaire (the part of the questionnaire related to the measurement with the rear camera of a smartphone) completed by each participant after the test procedure(i.e. after all measurements in the given participant).
The scale will be 1-5, with higher scores meaning better outcome.
day 1 (after a single visit, approximately 30 minutes)
See also
  Status Clinical Trial Phase
Recruiting NCT05563805 - Exploring Virtual Reality Adventure Training Exergaming N/A
Recruiting NCT04554160 - Arrhythmias in Pulmonary Hypertension Assessed by Continuous Long-term Cardiac Monitoring
Completed NCT02153385 - Effects of Yoga on Heart Rate Variability and Mood in Women N/A
Terminated NCT04381182 - Heart Rate Variability and Stress Management Enhancement N/A
Active, not recruiting NCT04342078 - Vitamin D and Health Outcomes in Preterm Born Population
Completed NCT03068182 - Effects of Arm Crank Versus Treadmill Exercises on Heart Rate Variability in Adult Subjects. N/A
Recruiting NCT06101225 - Effects of a Relaxation and Guided Imagery Intervention in School Context N/A
Recruiting NCT05679882 - Effects of Natural Sounds on Attention Restoration Outdoors N/A
Recruiting NCT05950698 - Obesity and Bariatric Surgery According to a Biopsychosocial Perspective.
Completed NCT05093439 - Heart Rate Variability at the Emergency Department
Completed NCT04487899 - Analysis of Heart Rate Variability During Emergency Flight Simulator Missions in Fighter Pilots
Recruiting NCT05560893 - The Heart of the Community Study N/A
Completed NCT04869176 - Effect of Caffeine on Heart Rate Variability in Newborns N/A
Recruiting NCT06425939 - Exploring the Relationship Between Heart Rate Variability (HRV), Training Load, and Exercise Performance N/A
Recruiting NCT05869721 - Effects of Yoga on Women With Breast Cancer N/A
Recruiting NCT06450483 - Compare the Effects of Different AVNS Parameters on Neck Pain, HRV, Balance in Collegiate Atheletes N/A
Completed NCT02137057 - The Effect of Spinal Anesthesia on Heart Rate Variability in Diabetic Patients N/A
Completed NCT01908920 - Osteopathic Manipulative Treatment Effects on High Frequency Parameters in Healthy Subjects N/A
Not yet recruiting NCT05296616 - Sleep Quality, Stress Level and Heart Rate Variability of University Students N/A
Recruiting NCT04826601 - The Physical and Psychologic Effects of Aromatherapy in Cancer Patients During Chemotherapy N/A