Clinical Trials Logo

Clinical Trial Summary

The purpose of the study is to perform the first clinical trial on human subjects using the Sonic Flashlight (SF) to guide placement of Peripherally Inserted Central Catheters (PICCs).


Clinical Trial Description

The SF prototype is an ultrasound (US) display device under development at our institution that replaces the traditional US monitor by using a half -silvered mirror to reflect real-time US images into the body. When the operator looks through the half -silvered mirror, the US image appears to float beneath the surface of the skin, exactly where it is being scanned. The SF merges the US image, probe, needle, operator's hands, and patient into the same field of view, making procedures more intuitive to the novice user, in contrast to conventional US (CUS) guided procedures, where hand-eye coordination is displaced, forcing the operator to look away from the field to see the US screen. We have previously shown that VA in phantoms is easier to use, faster to use, and faster to learn than CUS in novice and experienced US users. The current SF prototype uses a 10MHz US system (Terason, Burlington, MA) modified by attaching a small flat-panel display (AM550L OLED, Kodak, Rochester, NY) and a 20x50x1mm half -silvered mirror to the probe. A standard, clear, sterile polyethylene US probe cover was placed over the entire SF to sterilize the device. 15 subjects needing PICC lines were enrolled in this study. Scanning the upper arm of each patient with the SF, the basilic vein, brachial vein, and brachial artery were identified on the in situ US image. A 21ga needle was guided into the basilic or brachial vein. The number of attempts, or sticks, per PICC line placement, specific vein accessed, and reason for a failed attempt were recorded.

The vasculature was clearly visualized in situ using the SF. The needle was aimed and easily inserted into the basilic or brachial vein, and the needle tip visualized at its expected location. Successful vascular access was obtained with all 15 subjects, 13 of 15 on the first attempt, and 2 of 15 on the second attempt. 3 of the 15 lines encountered mild blurriness looking through the SF due to inconsistent optical properties of the probe cover.

This study shows that venous access can be obtained using the SF. Once the optical problems with the probe covers are addressed, the SF will be ready for a larger clinical trial. ;


Study Design

Allocation: Randomized, Endpoint Classification: Safety/Efficacy Study, Intervention Model: Single Group Assignment, Masking: Open Label, Primary Purpose: Treatment


Related Conditions & MeSH terms


NCT number NCT00226213
Study type Interventional
Source University of Pittsburgh
Contact
Status Completed
Phase Phase 1
Start date February 2005
Completion date June 2007

See also
  Status Clinical Trial Phase
Completed NCT00389558 - Antiseptic Use and Dressing Application Phase 4
Completed NCT01479153 - Venous Site for Central Catheterization N/A
Recruiting NCT01142934 - Trial on the Efficacy of Tegaderm Chlorhexidine Gluconate (CHG) in Reducing Catheter Related Bloodstream Infections Phase 4
Completed NCT00516906 - Clinical Performance of a Chlorhexidine Antimicrobial Dressing Phase 2/Phase 3
Not yet recruiting NCT03303274 - A Difference in Subclavian Vein Catheterization Between Supine and Lateral Tilt Position - Stage II N/A
Not yet recruiting NCT03296735 - A Difference in Subclavian Vein Catheterization Between Supine and Lateral Tilt Position - Stage I N/A
Completed NCT04714255 - Efficacy of Art Intervention on Decreasing Pain and Anxiety During Intravenous Cannulation N/A
Completed NCT03677765 - Comparisons of Complications Related to Two Approaches of Ultrasonography-guided Subclavian Venous Catheterization N/A
Recruiting NCT02299752 - Unnoticed Gloves Perforation N/A
Completed NCT00799981 - A Comparison of Two Urinary Catheters of Different Lengths for Female Use, in Intermittent Catheterization N/A
Completed NCT00802750 - Study Where Healthy Volunteers Evaluate LoFric Catheters N/A
Completed NCT03582540 - Early Versus Delayed Double-guidewire Technique in Difficult Biliary Cannulation. (DFG) N/A
Completed NCT02899546 - Study of Peripherally Inserted Central Catheter (PICC)-Related Infections in a Tropical Area
Completed NCT04865783 - Cryospray to Reduce Pain During Venous Cannulation N/A
Completed NCT02431858 - Catheter Over Needle vs Catheter Through Needle Phase 4
Completed NCT01133652 - Variation In Success of Intravenous (IV) Placement With Observation Using New Techniques Phase 4
Completed NCT05317533 - Prevalence of Distal Radial Artery Occlusion in Cardiovascular Catheterization Via Distal Radial Access
Completed NCT01347463 - Validation of the Ipsilateral Nipple as the Directional Guide for Internal Jugular Vein (IJV) Catheterization
Terminated NCT00482547 - Study of a Urethral Catheter Coated With Eluting Silver Salts (SUCCESS) Phase 4
Completed NCT00226226 - Peripherally Inserted Central Catheter Placement by IV Team Nurses Using the Sonic Flashlight Phase 1