Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT04392349
Other study ID # 1023707
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date July 1, 2020
Est. completion date December 10, 2020

Study information

Verified date February 2021
Source Haag-Streit AG
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The EYESTAR 900 with software version i9.5.1.0 includes new analysis functionality. Since its clinical performance cannot be assessed based solely on clinical literature as found in the Clinical Evaluation Report, further data from a clinical trial is required. The objective of this trial is to assess the clinical performance of the new features of the EYESTAR 900 with software version i9.5.1.0. This study is a necessary part of the clinical evaluation process of the investigational device. The results of this study are used for the clinical evaluation, and for reporting of in-vivo repeatabilities in the instructions for use of the investigational device, as required by topography standards.


Description:

Ocular biometry is the act of measuring the geometric properties of the eye, in particular distances between and thicknesses of the visual axis of the cornea, lens and retina, as well as curvature of the anterior cornea. These measurements are used mainly to determine implant type and dimensions for cataract surgery. Corneal topography is the act of measuring the shape of the cornea, in particular of the anterior cornea, but, depending on the device and the application, also of the posterior cornea, as well as the distance between the anterior to the posterior of the cornea. These measurements can be used in the context of ocular biometry, but are also useful for many other applications where knowledge of the optical and structural properties of the cornea are of interest. Optical coherence tomography (OCT) is a well-established imaging modality in ophthalmology. It uses interferometry to obtain a scattering profile of the eye along the direction of propagation of a laser beam which is directed onto the eye. In analogy to ultrasound imaging, this scattering profile is called A-Scan. By laterally translating the measurement beam, several A-scans can be combined to form a 2-dimensional image or 3-dimensional (3D) tomogram of the eye. The main use of OCT is the cross-sectional imaging of the retina or the cornea, primarily for diagnostic purposes[4]. Recently, anterior segment OCT has also been used for corneal topography[5], as well as for biometry and cross-sectional imaging along the entire length of the eye[6]. EYESTAR 900 is a device developed by Haag-Streit which utilises 3D OCT for quantitative measurements of the geometry of the entire eye, including ocular biometry and corneal topography. CE approval for EYESTAR 900 with software version i9.4.0.0 is pending and expected before the start of this clinical trial. The development of this device has been continued, and the following additions have been made in software version i9.5.1.0 used for this clinical study with respect to the software version i9.4.0.0: - Extended corneal topography: corneal topography can be evaluated over an extended area (diameter of trajectory d=13mm with respect to d=8.6mm in "standard corneal topography"). - Irregularity segmentation: segmentation methods (identification of corneal surfaces) used in corneal topography now incorporate an improved algorithm for segmentation of irregular corneal surfaces. - Irregularity visualization: visualization of topography maps has been supplemented by an overlay of regions with irregular corneal surfaces. - Zone-based keratometry: visualization of topography maps has been further supplemented by an overlay of local keratometry values (power and axes of flat and steep meridians in the centre, middle and outer corneal surface). - Crystalline lens tilt: the tilt of the crystalline lens normal vector with respect to the visual axis can be visualized and quantified. Both irregularity visualization and zone-based keratometry are applicable to corneal topography data acquired with the standard corneal topography protocol, as well as corneal topography acquired over the extended area (diameter 13mm, extended corneal topography). The primary objective of this clinical trial is to assess the clinical performance of extended corneal topography using irregularity segmentation, standard corneal topography using irregularity segmentation and crystalline lens tilt. To that end, for each measurand, the in-vivo repeatability will be quantified. For "standard corneal topography" and "extended corneal topography", the limits of agreement and the mean measurement deviation, with respect to three comparators are analyzed. As a secondary objective of the study, raw measurement data will be collected to allow for the improvement of existing algorithms, development of additional measurands and for retrospective analysis. Examinations with EYESTAR 900 and the other study devices are non-contact examinations. No diseases are studied as part of this study.


Recruitment information / eligibility

Status Completed
Enrollment 29
Est. completion date December 10, 2020
Est. primary completion date August 20, 2020
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: voluntary participation of subjects of legal age, irrespective of age, gender, ethnicity, of which: - participants with healthy eyes, - participants with corneal scarring, - participants with irregular astigmatism. Exclusion Criteria: - Involuntary participation or inability to provide consent, - persons of vulnerable populations, - persons unable or unwilling to follow instructions, - persons unable to maintain fixation for the duration of the examinations, - persons with active inflammation or infections of the eye, - persons with a tear film break up time of less than 5 sec.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Haag-Streit EYESTAR 900
The EYESTAR 900 is a new device developed by HAAG-STREIT AG which utilises Optical Coherence Tomography (OCT) for quantitative measurements of the geometry of the eye. As a successor to the LENSTAR 900, it represents the most recent in a series of successful biometry devices.
Tomey CASIA 2
The CASIA2 is a testing application for Cataract / Glaucoma / Cornea Surgery already established on the market. It is a representative device for measurement of anterior and posterior corneal curvature (anterior and posterior corneal topography) by optical coherence tomography, and for measurement of crystalline lens tilt by optical coherence tomography.
Oculus PENTACAM HR
The Pentacam HR is a high-resolution rotating Scheimpflug camera system for anterior segment analysis already established on the market. It is used as representative device for advanced anterior segment analysis such as anterior and posterior corneal topography and corneal pachymetry.
Zeiss ATLAS 9000
ATLAS is a corneal topography system used as representative device for measurement of anterior corneal curvature (anterior corneal topography).

Locations

Country Name City State
Switzerland University Hospital Basel Basel

Sponsors (1)

Lead Sponsor Collaborator
Haag-Streit AG

Country where clinical trial is conducted

Switzerland, 

Outcome

Type Measure Description Time frame Safety issue
Primary In-vivo Repeatability The in-vivo repeatability is calculated as the square root of the residual variance of a mixed-effects ANOVA model (using subject as a random effect, and eye within subject as a fixed effect), which we multiply with a factor of square root of two (in order to make it commensurate with the standard deviation of the differences between two independent measurements of each eye in the study sample).
For comparator devices, the in-vivo repeatability will be assessed as well, based on two independent measurements of the same eye.
3 months
Primary Limits of Agreement For each measurand, a comparator device was defined which represents the current clinical standard. For the comparator device, for each measured parameter the mean value per eye is obtained (this value serves as the "best guess" of the true value of this parameter for one particular eye). Then, for each examination of EYESTAR 900, the difference with respect to this comparator mean value is calculated. For the resulting differences per parameter, the interval spanning the mean difference ± 1.96 * the first standard deviation of the differences is obtained. This interval quantifies the limits of agreement, spanning 95% of the expected measurement differences between EYESTAR 900 and the comparator device. 3 months
Primary Confidence Interval of Differences The mean deviation with respect to a comparator device ("equivalence") is determined as the confidence interval of a double-sided Wilcoxon test (a=0.05) of the mean measurement difference between the investigational device and the comparator device, computed over the study sample. 3 months
Secondary Raw OCT Data Acquisition of volumetric OCT Data of the eye. This secondary study outcome allows for the improvement of existing algorithms, development of additional measurands and for retrospective analysis but will not be processed within the scope of this clinical trial. 3 months
Secondary Raw Image Data Acquisition of photographic images of the eye. This secondary study outcome allows for the improvement of existing algorithms, development of additional measurands and for retrospective analysis but will not be processed within the scope of this clinical trial. 3 months
See also
  Status Clinical Trial Phase
Completed NCT04685538 - Chloroprocaine 3% Gel Eye Drop as Topical Anestheticsin Phacoemulsification. Phase 3
Recruiting NCT06060041 - IC-8 Apthera IOL New Enrollment Post Approval Study
Recruiting NCT05518539 - Evaluation of Quality of Vision and Visual Outcomes With Bilateral Implantation of the Clareon PanOptix Intraocular Lens
Recruiting NCT05271942 - Tilt and Tumble vs Divide and Conquer - a Unique Comparison of the Two Cataract Surgery Methods N/A
Active, not recruiting NCT04778501 - PMCF Study on Monofocal Toric IOL (PODEYE TORIC) in Asia N/A
Completed NCT05062564 - Efficacy of LipiFlow in Patients Affected by Meibomian Gland Dysfunction in Reducing Post-cataract Surgery Dry Eye N/A
Completed NCT03751033 - Influence of DisCoVisc Ophthalmic Viscosurgical Device (OVD) on Intraoperative Aberrometry Readings N/A
Completed NCT02529488 - Investigation of AcrySof® IQ PanOptix™ Presbyopia-Correcting Intraocular Lens (IOL) Model TFNT00 N/A
Completed NCT04539548 - A Study Assessing the Safety and Efficacy of Dextenza® for the Treatment of Ocular Pain and Inflammation Following Surgery for Pediatric Cataract Phase 3
Completed NCT03740659 - Evaluation Of Aqueous Humor Of Levofloxacin-Dexamethasone Eye Drops And Of Its Components In Patients Undergoing Cataract Surgery Phase 2
Completed NCT03494257 - Effect of Fixed Brinzolamide-brimonidine Combination on Intraocular Pressure After Phacoemulsification N/A
Completed NCT05119127 - Rotational Stability of Acrysof IQ Vivity Extended Vision Toric IOL and Refractive Visual Outcome. N/A
Active, not recruiting NCT04271709 - Manhattan Vision Screening and Follow-Up Study (NYC-SIGHT) N/A
Recruiting NCT03713268 - Intraoperative OCT Guidance of Intraocular Surgery II
Completed NCT03739528 - Levo-Dexa vs. Tobra+Dexa for Prevention and Treatment of Inflammation and Prevention of Infection in Cataract Surgery Phase 3
Completed NCT02888210 - A Study Assessing Safety and Efficacy of MD-15 Intraocular Lens in Patients With Aphakic Eye After Cataract Surgery Phase 3
Completed NCT03356847 - Evaluation of the Rotational Stability of the Monofocal SISA Implant Following Cataract Surgery N/A
Completed NCT04332640 - Clinical Evaluation of the Next Generation Phaco System N/A
Recruiting NCT03638726 - Subconjunctival Atropine and Intracameral Epinephrine for Pupil Dilation in Phacoemulsification Phase 4
Completed NCT03050697 - Evaluation of the Safety and Performance of the HARMONI® Toric Lens N/A