Clinical Trials Logo

Clinical Trial Summary

Abstract Purpose: To evaluate the visual performance and rotational stability of the Tecnis Toric one-piece IOL (AMO, USA) during the first 3 post-operative months Design: Prospective single center study Setting: VIROS - Vienna Institute for Research in Ocular Surgery, a Karl Landsteiner Institute, Hanusch Hospitel, Vienna, Austria Methods: In this prospective study patients with age-related cataract and corneal astigmatism of 1.0 to 3.0 D measured with the IOL-Master 500 (Carl Zeiss Meditec AG, Germany) were included. Pre-operatively, rotating Scheimpflug scans (Pentacam HR, Oculus, Germany) were performed and the cornea was marked in the sitting position at the slit lamp. Patients recieved a single-piece toric hydrophobic acrylic IOL (Tecnis Toric, AMO, USA). Immediately and 3 months after surgery retroillumination photographs were taken to assess the rotational stability of the IOL. Additionally, Autorefraction (Topcon, USA), subjective refraction, uncorrected and distance corrected visual acuity, keratometry, Scheimpflug and ocular wavefront (WASCA, Carl Zeiss Meditec AG, Germany) measurements were performed at the 3 months follow-up.


Clinical Trial Description

Whereas first designs of toric intraocular lenses (toric IOLs) in the early 1990's showed an IOL rotation of more than 30° in one fifth of the patients, modern toric IOLs typically show a mean absolute rotation of 3° to 5°, which would result in a loss of about 10% to 15% of the astigmatism reducing effect of the toric IOL.

Rotational stability of a toric IOL depends on the interaction between the toric IOL and the posterior capsule, whereas misalignment of the toric IOL (defined in this study as the difference between the 3 months postoperatively measured axis of the toric IOL and the intended axis) depends on several factors additionally to rotational stability. Intraoperatively, misalignment may happen due to cyclotorsion of the eye in the lying position or due to peribulbar anaesthesia, and due to imprecision of the surgeon when positioning of the IOL relative to the intended meridian. Both these imprecisions can be dealt with by pre-operative marking of the eye in the sitting position and diligence by the surgeon. Postoperatively, the IOL may rotate because it is undersized for the capsule bag or due to the capsule shrinkage that takes place during fibrotic contraction of the bag in the postoperative period. Since most current IOLs are slightly oversized for the capsule bag, the former is observed rarely and would be more likely in long eyes which tend to have a larger capsule bag diameter. However, capsule bag shrinkage is thought to induce rotation in IOLs with open-loop haptics due to the asymmetry of the haptic design. Typical IOL haptic designs that improve rotational stability are either plate haptic IOLs or special Z-haptic shaped open-loop haptics that attempt to counteract the rotational effect of compression of the shrinking bag. However, both of these IOL designs are thought to have downsides. The former seems to have an increased risk of posterior capsule opacification due to a less effective lens epithelial barrier effect of the optic edge and may also show more rotation in the bag immediately after surgery due to the shorter haptic overall length. The latter IOL design, such as the Z-haptic IOL, is cumbersome to implant and may be prone to damage during implantation.

Aim of this study was to evaluate the rotational stability of a novel single-piece hydrophobic acrylic toric IOL with a C-haptic design. ;


Study Design

Observational Model: Cohort, Time Perspective: Prospective


Related Conditions & MeSH terms


NCT number NCT02155959
Study type Observational
Source Vienna Institute for Research in Ocular Surgery
Contact
Status Completed
Phase N/A
Start date March 2011
Completion date December 2012

See also
  Status Clinical Trial Phase
Completed NCT04685538 - Chloroprocaine 3% Gel Eye Drop as Topical Anestheticsin Phacoemulsification. Phase 3
Recruiting NCT06060041 - IC-8 Apthera IOL New Enrollment Post Approval Study
Recruiting NCT05518539 - Evaluation of Quality of Vision and Visual Outcomes With Bilateral Implantation of the Clareon PanOptix Intraocular Lens
Recruiting NCT05271942 - Tilt and Tumble vs Divide and Conquer - a Unique Comparison of the Two Cataract Surgery Methods N/A
Active, not recruiting NCT04778501 - PMCF Study on Monofocal Toric IOL (PODEYE TORIC) in Asia N/A
Completed NCT05062564 - Efficacy of LipiFlow in Patients Affected by Meibomian Gland Dysfunction in Reducing Post-cataract Surgery Dry Eye N/A
Completed NCT03751033 - Influence of DisCoVisc Ophthalmic Viscosurgical Device (OVD) on Intraoperative Aberrometry Readings N/A
Completed NCT02529488 - Investigation of AcrySof® IQ PanOptix™ Presbyopia-Correcting Intraocular Lens (IOL) Model TFNT00 N/A
Completed NCT04539548 - A Study Assessing the Safety and Efficacy of Dextenza® for the Treatment of Ocular Pain and Inflammation Following Surgery for Pediatric Cataract Phase 3
Completed NCT03740659 - Evaluation Of Aqueous Humor Of Levofloxacin-Dexamethasone Eye Drops And Of Its Components In Patients Undergoing Cataract Surgery Phase 2
Completed NCT03494257 - Effect of Fixed Brinzolamide-brimonidine Combination on Intraocular Pressure After Phacoemulsification N/A
Completed NCT05119127 - Rotational Stability of Acrysof IQ Vivity Extended Vision Toric IOL and Refractive Visual Outcome. N/A
Active, not recruiting NCT04271709 - Manhattan Vision Screening and Follow-Up Study (NYC-SIGHT) N/A
Recruiting NCT03713268 - Intraoperative OCT Guidance of Intraocular Surgery II
Completed NCT03739528 - Levo-Dexa vs. Tobra+Dexa for Prevention and Treatment of Inflammation and Prevention of Infection in Cataract Surgery Phase 3
Completed NCT02888210 - A Study Assessing Safety and Efficacy of MD-15 Intraocular Lens in Patients With Aphakic Eye After Cataract Surgery Phase 3
Completed NCT03356847 - Evaluation of the Rotational Stability of the Monofocal SISA Implant Following Cataract Surgery N/A
Completed NCT04332640 - Clinical Evaluation of the Next Generation Phaco System N/A
Recruiting NCT03638726 - Subconjunctival Atropine and Intracameral Epinephrine for Pupil Dilation in Phacoemulsification Phase 4
Completed NCT03050697 - Evaluation of the Safety and Performance of the HARMONI® Toric Lens N/A