Bradycardia Clinical Trial
— TEMPACEOfficial title:
Safety and Effectiveness of the External Fastening System for Temporary Active Fixation Pacemakers (TEMPACE)
Temporary cardiac pacing with active-fixation leads (TPAFL) using a reusable permanent pacemaker generator has been shown to be safer than lead systems without fixation. However, TPAFL requires the off-label (non-certificated) use of Pacemaker leads and generators. It is designed a fastening system to ensure the safety and efficacy of the procedure: the KronoSafe System®. The aim of this study is to demonstrate the safety and effectiveness of the KronoSafe System® for temporary pacing in a series of patients receiving TPAFL.
Status | Recruiting |
Enrollment | 122 |
Est. completion date | October 1, 2024 |
Est. primary completion date | October 1, 2024 |
Accepts healthy volunteers | No |
Gender | All |
Age group | 18 Years and older |
Eligibility | Inclusion Criteria: - Patient older than 18 years. - Need for temporary cardiac stimulation - Temporary active fixation pacemaker implantation through left or right jugular access. - Signature of informed consent - Correct functioning of the electrode after the implantation: sensing of the R wave above 5 millivolts. Ventricular pacing threshold below 1 volt. Electrode impedance between 400-1200 Ohm. - Pacemaker programming with stimulation output at 5 volts intensity and 1 millisecond pulse duration. - Placement of the KronoSafe® fastening system after electrode implantation. Exclusion Criteria: - Complication during the procedure: pneumothorax, hematoma at the puncture site, cardiac perforation, accidental arterial puncture or other that, in the investigator's opinion, poses an added risk of pacemaker dysfunction. - Need to transfer the patient to another center with loss of follow-up. - Absence of surveillance through continuous electrocardiographic monitoring with detection of arrhythmic events. |
Country | Name | City | State |
---|---|---|---|
Spain | Hospital General Universitario de Alicante Dr.Balmis | Alicante | |
Spain | Hospital Universitari Vall D'Hebron | Barcelona |
Lead Sponsor | Collaborator |
---|---|
ICU Medical Technologies S.L. | Universidad Miguel Hernandez de Elche |
Spain,
Betts TR. Regional survey of temporary transvenous pacing procedures and complications. Postgrad Med J. 2003 Aug;79(934):463-5. doi: 10.1136/pmj.79.934.463. — View Citation
Kawata H, Pretorius V, Phan H, Mulpuru S, Gadiyaram V, Patel J, Steltzner D, Krummen D, Feld G, Birgersdotter-Green U. Utility and safety of temporary pacing using active fixation leads and externalized re-usable permanent pacemakers after lead extraction — View Citation
Vicente-Miralles R, Martin-Langerwerf DA, Nunez-Martinez JM, Marco-Juan A, Sangio-Ferreyros AD, Davila-Arias A, Montes-Jacobo L, Barrios V, Palazon-Bru A, Gil-Guillen VF, Bertomeu-Gonzalez V. A new fastening system for temporary pacing with active-fixatio — View Citation
Zei PC, Eckart RE, Epstein LM. Modified temporary cardiac pacing using transvenous active fixation leads and external re-sterilized pulse generators. J Am Coll Cardiol. 2006 Apr 4;47(7):1487-9. doi: 10.1016/j.jacc.2006.01.006. Epub 2006 Mar 15. No abstrac — View Citation
Type | Measure | Description | Time frame | Safety issue |
---|---|---|---|---|
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | Immediately after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | Immediately after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | Immediately after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | 24 hours after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 3 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 5 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 7 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 9 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 11 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 13 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 15 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 17 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 19 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 21 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 23 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 25 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 27 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | day 29 after the procedure of lead implantation | |
Primary | Ventricular R wave detection | Ventricular R wave, measured in millivolts with pacemaker generator programmer | Immediately before the procedure of lead removal | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | 24 hours after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 3 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 5 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 7 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 9 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 11 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 13 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 15 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 17 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 19 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 21 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 23 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 25 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 27 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | day 29 after the procedure of lead implantation | |
Primary | Lead impedance detection | Lead impedance, measured in ohms with pacemaker generator programmer | Immediately before the procedure of lead removal | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | 24 hours after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 3 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 5 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 7 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 9 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 11 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 13 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 15 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 17 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 19 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 21 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 23 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 25 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 27 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | day 29 after the procedure of lead implantation | |
Primary | Right ventricle threshold detection | Right ventricle threshold, measured in volts with pacemaker generator programmer | Immediately before the procedure of lead removal | |
Primary | Lead displacement detection | Visual observation of 2 cm electrode displacement through vein acces by clinical examination | Immediately before the procedure of lead removal | |
Primary | Capture failure detection | Right ventricle threshold above 5 volts during 1 millisecond, measured with pacemaker generator programmer. | Immediately before the procedure of lead removal | |
Primary | Sensing failure detection | Ventricular R wave below 1 volt, measured with pacemaker generator programmer | Immediately before the procedure of lead removal | |
Primary | Lead insulation breakdown detection | Visual observation of lead insulation breakdown by clinical examination | Immediately before the procedure of lead removal | |
Primary | Generator detachment detection | Visual observation of generator detachment by clinical examination. removal of the generator from the casing. | Immediately before the procedure of lead removal | |
Primary | Cardiac peroration detection | New pericardial effusion detected by Echocardiography with signs of cardiac tamponade: respiratory variation of transmitral filling flow greater than 25% | Immediately before the procedure of lead removal | |
Secondary | Detection of Bacteremia | Isolation of the same bacteria in the pacemaker lead and 2 blood cultures from the patient | Immediately after the procedure of lead removal | |
Secondary | Detection of Agitation | Record of antipsychotic and anxiolytic treatment of the patient | Immediately after the procedure of lead removal |
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT03922386 -
Safety and Electrical Performances of XFINE Leads
|
N/A | |
Completed |
NCT03294018 -
Heart Rate Changes Following the Administration of Sugammadex
|
||
Not yet recruiting |
NCT05649891 -
Checklists Resuscitation Emergency Department
|
N/A | |
Not yet recruiting |
NCT05440071 -
PoSt-market Assessment on Safety and Performance of J-sHAPed xFINE Lead
|
||
Completed |
NCT02153242 -
The Effects of Selective Site Right Ventricular Pacing
|
||
Recruiting |
NCT01825473 -
Study of Erythromycin in GER-Associated Apnea of the Newborn
|
N/A | |
Completed |
NCT00950287 -
Detection of Neonatal Bradycardia
|
N/A | |
Completed |
NCT00721136 -
Randomized Study of the Use of Warfarin During Pacemaker or ICD Implantation in Patients Requiring Long Term Anticoagulation
|
N/A | |
Completed |
NCT00286858 -
Beluga - Clinical Observations of Automatic Algorithms for Cardiac Pacing
|
||
Completed |
NCT00158925 -
The EASYTRAK EPI Clinical Investigation
|
Phase 3 | |
Completed |
NCT00180557 -
Austria Study - Analysis of Difference Between Active and Passive Fixation Leads
|
Phase 4 | |
Completed |
NCT01076348 -
Model 4965 Post-Approval Study
|
||
Completed |
NCT04198220 -
BIO|STREAM.ICM Obesity
|
||
Recruiting |
NCT05298748 -
The Effect of Womb Recordings on Maturation of Respiratory Control in Preterm Infants
|
N/A | |
Recruiting |
NCT04353960 -
The Alaska Oculocardiac Reflex Study
|
||
Terminated |
NCT04093414 -
Left Bundle Area Versus Selective His Bundle Pacing
|
N/A | |
Completed |
NCT03786640 -
Abbott Brady 3T MRI PMCF
|
||
Recruiting |
NCT05932602 -
AVEIR DR Coverage With Evidence Development (CED) Study
|
||
Recruiting |
NCT05935007 -
Aveir DR Real-World Evidence Post-Approval Study
|
||
Recruiting |
NCT04075084 -
Observation of Clinical Routine Care for Patients With BIOTRONIK Implantable Cardiac Monitors (ICMs)
|