Clinical Trials Logo

Clinical Trial Summary

Purpose To characterize the safety and clinical benefit of the Engager Transcatheter Aortic Valve Implantation System in a real world environment.

Design Prospective non-randomized, non-interventional, multicenter, single-arm, post-market study.

250 Patients will have the Engager bioprosthesis implanted. Patients will be followed at 1 month, 6 months, and 12 months after the procedure. The recruitment period will be approximately 12 months, so the total study duration will be 24 months.


Clinical Trial Description

The study objective is to characterize the safety and clinical benefit of the Engager Transcatheter Aortic Valve Implantation System in a real world setting.

The primary endpoint of this study is all-cause mortality at 30 days or during index hospitalization.

Secondary endpoints of this study are:

- Device success according to VARC2 is defined as:

- absence of procedural mortality (all-cause mortality at discharge, 30 days or during index hospitalization), AND

- correct positioning of the Engager bioprosthesis in the proper anatomic location, AND

- Engager bioprosthesis performing as intended (no prosthesis-patient mismatch and mean aortic valve gradient < 20 mmHg or peak velocity < 3 m/s, AND no moderate or severe regurgitation).

- Composite early safety (at 30 days) according to VARC2 includes the following components:

- all-cause mortality

- all stroke

- life-threatening bleeding

- acute kidney injury (stage 2-3)

- coronary artery obstruction requiring intervention

- major vascular complication

- valve-related dysfunction requiring repeat procedure (BAV, TAVI or SAVR)

- Clinical efficacy according to VARC2 at 6 months and 1 year is defined as:

- all-cause mortality

- all stroke

- hospitalizations for valve-related symptoms or worsening congestive heart failure (CHF)

- NYHA class III or IV

- valve-related dysfunction (mean aortic valve gradient ≥20 mm Hg, effective orifice area (EOA) ≤0.9-1.1 cm² and/or Doppler velocity index (DVI) <0.35 m/s, AND/OR moderate or severe prosthetic valve regurgitation)

- Time-related valve safety accordingly to VARC2 at 6 months and 1 year is defined as:

- Structural valve deterioration

- Valve-related dysfunction (mean aortic valve gradient ≥20 mm Hg, (EOA) ≤0.9-1.1 cm² and/or (DVI) <0.35 m/s, AND/OR moderate or severe prosthetic valve regurgitation)

- Requiring repeat procedure (TAVI or SAVR)

- Prosthetic valve endocarditis

- Prosthetic valve thrombosis

- Thrombo-embolic events (e.g. stroke)

- VARC bleeding, unless clearly unrelated to valve therapy (e.g. trauma)

- All-cause mortality at 6 months and 1 year

- Cardiovascular mortality at 30 days, 6 months and 1 year

- Incidence of TAVI-related complications at 30 days, 6 months and 1 year:

- Myocardial infarction

- Periprocedural and spontaneous stroke

- Life-threatening, major and minor bleedings

- Acute kidney injury state 1, 2 and 3

- Apical and vascular complications

- Conduction disturbances

- Conversion to open-heart surgery

- Unplanned use of cardiopulmonary bypass

- Coronary obstruction and dissection

- Ventricular septal perforation

- Mitral valve apparatus damage or dysfunction

- Cardiac tamponade

- Endocarditis

- Valve thrombus

- Valve migration, embolization or ectopic valve deployment

- Valve in valve deployment

- Total, transvalvular and paravalvular regurgitation as measured by TTE at 30 days and 1 year

- Mean aortic valve gradient, peak velocity and effective orifice area as measured by TTE at 30 days and 1 year

- Engager implant depth as measured by standard of care angiography at the end of implant procedure

- Incidence and rationale for permanent pacemaker implantation after Engager implant procedure

- Changes in Quality of Life as measured by the EQ5D questionnaire at baseline, discharge, 30 days, 6 months and 1 year

- Length of stay during index hospitalization (ICU and general ward)

- Incidence of re-hospitalization at 30 days, 6 months and 1 year

Subject population All patients with severe aortic stenosis requiring treatment, who are selected by a heart team to be treated electively with the Engager Transcatheter Aortic Valve Implantation System are eligible to be included in this study.

Treatment The Engager Transcatheter Aortic Bioprosthesis will be implanted using the Engager Transcatheter Delivery System. The bioprosthesis is delivered transapically on a catheter during a beating heart procedure. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT02149654
Study type Observational [Patient Registry]
Source Medtronic Cardiovascular
Contact
Status Completed
Phase
Start date May 2014
Completion date January 2017

See also
  Status Clinical Trial Phase
Completed NCT03186339 - Validation of the "TASQ" in Patients Undergoing SAVR or TF-TAVI
Recruiting NCT03549559 - Imaging Histone Deacetylase in the Heart N/A
Terminated NCT02854319 - REpositionable Percutaneous Replacement of NatIve StEnotic Aortic Valve Through Implantation of LOTUS EDGE Valve System N/A
Recruiting NCT05601453 - The ReTAVI Prospective Observational Registry
Withdrawn NCT05481814 - CPX in Paradoxical Low Flow Aortic Stenosis
Completed NCT02241109 - Predicting Aortic Stenosis Progression by Measuring Serum Calcification Propensity N/A
Completed NCT01700439 - Surgical Treatment of Aortic Stenosis With a Next Generation, Rapid Deployment Surgical Aortic Valve N/A
Recruiting NCT04429035 - SLOW-Slower Progress of caLcificatiOn With Vitamin K2 N/A
Completed NCT04103931 - Impact of a Patient Decision Aid for Treatment of Aortic Stenosis N/A
Completed NCT03950440 - Assessing the Incidence of Postoperative Delirium Following Aortic Valve Replacement
Active, not recruiting NCT02661451 - Transcatheter Aortic Valve Replacement to UNload the Left Ventricle in Patients With ADvanced Heart Failure (TAVR UNLOAD) N/A
Completed NCT02847546 - Evaluation of the BARD® True™ Flow Valvuloplasty Perfusion Catheter for Aortic Valve Dilatation N/A
Completed NCT02792452 - Clinical Value of Stress Echocardiography in Moderate Aortic Stenosis
Completed NCT02758964 - Evaluation of Cerebral Thrombembolism After TAVR
Not yet recruiting NCT02536703 - Safety and Efficacy of Lotus Valve For TAVI In Patients With Severe Aortic Stenosis In Chinese Population Phase 3
Not yet recruiting NCT02541877 - Sizing-sTrategy of Bicuspid AoRtic Valve Stenosis With Transcatheter Self-expandable Valve Phase 3
Not yet recruiting NCT02221921 - Safety and Efficacy Study of MicroPort's Transcatheter Aortic Valve and Delivery System for TAVI N/A
Completed NCT02249000 - BIOVALVE - I / II Clincial Investigation N/A
Active, not recruiting NCT02080299 - Protection by Remote Ischemic Preconditioning During Transcatheter Aortic Valve Implantation Phase 2
Terminated NCT01939678 - Characterization and Role of Mutations in Sodium-phosphate Cotransporters in Patients With Calcific Aortic Valve Disease