Anesthesia Clinical Trial
Official title:
Comparison of the Efficiency of Anesthetic Depth Control Using Closed Loop Anesthesia Delivery System vs. Manual Control Using Target Controlled Infusion in Patients With Moderate to Severe Left Ventricular Systolic Dysfunction
Advancement in techniques for anaesthetic drug delivery and real time monitoring has
facilitated safe induction and maintenance of anaesthesia in severely compromised patients.
Cardiac diseases are the commonest causes of morbidity and left ventricular failure is the
commonest clinical presentation at the end stage. LV systolic dysfunction is defined as
reduction in LVEF ≤55%. Patients with LVEF 55%-46% have mild, 45%-36% moderate and ≤35%
severe LV systolic dysfunction. Patients with heart failure have a diminished cardiac
reserve capacity that may be further compromised by anaesthesia. In addition to depression
of sympathetic activity, most anaesthetics interfere with cardiovascular performance, either
by a direct myocardial depression or by modifying cardiovascular control mechanisms.
Propofol with fentanyl is advocated as the best anaesthetic combination for induction of
anaesthesia in patients undergoing CABG. Propofol is a drug with narrow therapeutic index
and may cause severe hypotension and hemodynamic instability during induction of
anaesthesia, especially if it is given in too large doses.
Automated drug delivery systems are popular for delivery of propofol. They can be of two
types, depending on whether they are based on pharmacokinetic or pharmacodynamic principles.
Closed Loop Anaesthesia Delivery system has been used world-wide and in our institute in
patients of various age groups and in patients undergoing cardiac surgery. But still the
studies are lacking in patients with moderate to severe left ventricular systolic
dysfunction. Moreover none of the studies have compared the efficacy of anaesthetic drug
delivery using these two devices in this group of patients.
Thus there is paucity of literature regarding PK and PD of propofol in patients with cardiac
failure. The investigators hypothesized that as the Closed Loop Anaesthesia Delivery System
is based on pharmacodyanamic principles, it should perform better than the Target Control
Infusion system, which works on pharmacokinetic principles. The investigators planned to
conduct this study to determine the anaesthetic depth control using Closed Loop Anaesthesia
Delivery system vs. manual control using Target Controlled Infusion in patients with
moderate to severe left ventricular systolic dysfunction.
MATERIALS AND METHODS:
Patient will be pre-medicated with oral tablet alprazolam 0.25 mg the night before and on
the morning of surgery. In the anesthetic room, 16 G i.v. cannula will be inserted and
following monitors will be attached: continuous pulse-oximetry (SpO2), electrocardiogram
(ECG), periodic non-invasive blood pressure (S/Anesthesia monitor, Datex Ohmeda Inc.,
Madison, WI) , and continuous BIS (BIS XP, Aspect Medical Systems, Newton, MA in the S/5
Anesthesia monitor). Pre-induction arterial line and central line will be inserted in all
patients and continuous arterial blood pressure and central venous pressure measurements
will be recorded. Pulmonary artery catheter will also be inserted in all patients for
monitoring of cardiac output. All invasive lines will be inserted under local anesthesia.
NIBP measurements will be stopped after transducing arterial line, because it may interfere
with the assessment of MOAA/S score.
All the patients will receive i.v. fentanyl at a dose of 3µg.kg-1 over a period 3 minutes,
followed by propofol administration either by CLADS or TCI.
Group 1:
Propofol will be administered with the Diprifusor (Marsh Pharmacokinetic model, Master TCI
pump, Fresenius Kabi; Bad Homburg; Germany) starting at a predetermined target plasma
propofol concentration (Cp) of 1.8µg.ml-1 based on a previous study. Further target
concentrations were set according to the Dixon up and down method. A failure will be
followed by an increase in the target plasma concentration in the next patient by 0.2µg.ml-1
and a success will be followed by a decrease in the target plasma concentration in the next
patient by 0.2µg.ml-1. TCI pump will be controlled manually and all the data including BIS
and vitals will be recorded through CLADS into a laptop. MOAA/S will be assessed every 15
second for assessing time to loss of consciousness; BIS and HR will be recorded every 5
seconds and MAP and PAP every 10 seconds. Thus, the anesthetic depth control will be
assessed based on time to achieve loss of consciousness and by comparing the dose of
propofol required for induction and maintenance. BIS overshoots and hemodynamic stability
will be noted. In case the patient achieves loss of consciousness within 5 minutes, a 2
minutes observation period will be allowed for achieving target BIS. If the patient does not
achieve target BIS after this period, the case will be considered a failure. On the other
hand, if the patient achieves target BIS within this period it will be considered a success.
In cases where neither LOC nor target BIS will be achieved within 5 minutes will also be
considered a failure Vecuronium will be administered at a dose of 0.1 mg.kg-1 after the LOC
and the patient will be intubated after 4 minutes. Patients will be ventilated through face
mask to maintain normocapnia administering air-oxygen mixture with FiO2 of 0.6 during the
period of induction till intubation. Fentanyl infusion 1µg.kg.h-1 will provide analgesia
with additional boluses of 1µg.kg-1 before skin incision, sternotomy and at the commencement
of CPB. Additional bolus of analgesic supplements (fentanyl 1µg.kg-1) will be given when the
mean arterial pressure (MAP) or heart rate exceeds 25% of the baseline <50. If hypertension
or tachycardia persists with a BIS≤50, either nitroglycerine infusion or esmolol will be
used. In conditions of hypotension, inotropic support and/or vasopressor will be initiated
after ensuring normovolaemia. Atropine sulphate will be used to treat bradycardia (heart
rate <45b.p.m.) after excluding other treatable causes. The MAP on CPB will be maintained
between 50 and 80 mmHg using phenylephrine / nitroglycerine as required. Systemic
hypothermia up to 28ºC will be practiced during CPB and patients will be actively re-warmed
to 36ºC before separation from CPB.
The number of patients exhibiting episodes of hypotension or hypertension and the number of
episodes of hypotension or hypertension requiring a change in the propofol target
concentration or the administration of any vasoactive drugs will be recorded.
Based on response of the previous patient to propofol administration using Diprifusor at a
predetermined target plasma propofol concentration (Cp), further increments or decrements of
Cp in the subsequent patients will be based on Dixon's up and down method. The target Cp
will be increased by 0.2 µg.ml-1 in patients who do not achieve BIS of 50 and Cp will be is
decreased in subsequent cases by 0.2µg.ml-1 in patients who do. Thus, six to eight pairs of
failure to success transition will be needed for calculation of the EC50 (i.e. the target
concentration at which 50% of the patients achieve BIS of 50). The mean of the mid value of
these transitions will be used to derive the EC50.
At the end of the study the BIS data will be analyzed for offline determination of induction
time. It is defined as the time at which the BIS reaches ≤50 for 2 consecutive readings from
the start of propofol infusion.
Group 2:
In the CLADS group induction and maintenance will be carried out automatically according to
the BIS. Hemodynamics will be controlled automatically through CLADS as well as through
manual ionotropic support as described above.
CLADS is a pharmacodynamic-pharmacokinetic model-based adaptive infusion system, which uses
BIS as the controlled variable and standard infusion pump as the 'actuator'. The 'control
algorithm' is based on the relation between various rates of propofol infusion (producing
different plasma concentrations) and BIS, taking into consideration the pharmacokinetic
variables (distribution, clearance). This was established in the developmental stage of
CLADS. The algorithm alters the rate of propofol infusion to steer and maintain BIS to the
set target taking into account existing BIS, time-elapsed since the initiation of infusion,
pharmacokinetics, time-delay factor between sensing and averaging of BIS data, time-delay
factor between change in infusion rate and actual change in the plasma concentration of
propofol, as well as the peak effect of propofol.
An IBM-compatible PC with PENTIUM 4 or higher processor is used to implement the control
algorithm, provide a user interface, and control communication through serial ports (RS 232)
with infusion system (Pilot-C; Fresenius, Paris, France) and vital sign monitor (AS5; Datex
Ohmeda Division, GE Healthcare, Singapore). The drug delivery can operate in two modes
—manual and automatic. In manual mode, rate of propofol infusion is controlled manually
through the keyboard. In 'automatic' mode, the system automatically controls propofol /
isoflurane administration as per the control algorithm. The automatic mode further has three
options: (i) induction, (ii) maintenance, and (iii) induction combined with maintenance. The
user must enter a target BIS value, age, weight, height along with the status of patient—low
risk (ASA I-III), high risk (ASA III-IV, NYHA class III), very high risk (ASA IV-V, NYHA
III-IV), or pediatric. The system can also function in 'monitor' mode, where it only updates
BIS and other patient data and provides a graphic display of current and trend values.
The system updates the electroencephalographic data every 5 s and calculates the BIS error
(target BIS—actual BIS). It uses PID (proportional integral differential) algorithm based on
this error to make the changes in propofol infusion rate to achieve target BIS. The
algorithm fine tunes the rate and duration of propofol delivery differently during induction
and maintenance phases of anesthesia delivery. During induction, controller tries to achieve
the target concentration in a stepwise fashion (while continuously receiving feedback of BIS
every 5 s) on the basis of the relationship between plasma concentrations and BIS. During
maintenance, 30 s is deemed as one epoch. The initial three as well as the last three BIS
values of each epoch are averaged and compared to assess the trend. When the trends indicate
an increasing BIS, higher target concentrations and so higher propofol rates are set and
vice versa if the trends indicate a decreasing BIS. These trends are also cross-checked with
larger epoch trends before making drug alterations.
A safety feature has been added wherein propofol and other anesthetics are stopped
automatically whenever hemodynamics drops below the safety limits set by the operator. This
would restart automatically when hemodynamics improves to values above the predefined lower
limit. The time delay for this automatic cut-off is at the most 10 s, which is the interval
at which the vitals are updated in the controller.
Time to achieve loss of consciousness, dose of propofol required for induction and
maintenance, BIS overshoots and hemodynamics will be recorded in patients administered
propofol using CLADS.
At the end of the surgery, CLADS/TCI will be discontinued and patient will be shifted to the
post-surgical intensive care unit for elective mechanical ventilation.
STATISTICAL ANALYSIS- Physiologic data are presented as mean (SD) and time intervals are
presented as median (range). The performance of the system was assessed by calculating
median performance error (MDPE), median absolute performance error (MDAPE), wobble and
divergence (time related trends) using methods of Varvel et al.
The performance error is given by the formula:
PE = [(BISmeasured-BIStarget)÷BIStarget]×100 The median performance error (MDPE) which
reflects the bias of CLADS in the ith subject is determined as MDPE=median {PEij
j=1........Ni } where Ni is the number of (PE) values obtained in the ith sub Median
Performance Error (MDPE) Median Absolute Performance Error (MDAPE)= median {I PEij
j=1........Ni I} Wobble is determined from the equation Wobble=median {I PEij - MDPEiI ,
j=1......Ni }
Global Score (GS) - The Global Score is used to score EEG-guided depth of anesthesia control
systems with one scalar:
GS=(MDAPE+WOBBLE)÷fraction of time BIS∈(40,60) Divergence is the slope of the linear
regression equation of absolute performance error against time and is expressed in units of
percentage divergence per minute. A positive value indicates progressive widening of the gap
between targeted and measured values, whereas a negative value indicates that the measured
values are converging on target. MDPE and MDAPE are measures of bias and precision,
respectively, wobble measures the intra- individual variability in the performance errors.
The percent of time when BIS remains within ±10 of target BIS during closed-loop (CLADS
group) or manual control (TCI group) will also be calculated. Adequacy of hemodynamic
control will be adjudged by the percentage of anesthesia time the mean arterial blood
pressure and heart rate (HR) are within ±25% of baseline.
Differences between the groups will be analyzed using the unpaired t-test for parametric
data and the Mann- Whitney U test for non-parametric data. Count data will be analysed using
the chi-square test. All analyses will be performed using SPSS v21.0 for Windows (SPSS Inc.,
Chicago IL, U.S.A.) and a P value <0.05 will be taken to be significant.
For calculation of the EC50 (plasma concentration at which 50% patients achieve a BIS) of
propofol required for induction will be calculated using Dixon and Massey method. Six to
eight pairs are necessary to calculate EC50 from Dixon up and down method. Probit analysis
of the data for TCI group will be then performed to determine EC50 and EC95.
From a pilot study, the investigators estimated that manual control using TCI maintained BIS
in the range of ±10 of the target 70% of the time. In order to assess the 20% improvement
from this control, the investigators calculated that they would need to recruit 36 patients
to achieve 80% power at 5% Type I error. A total of 40 patients were recruited, taking into
account possible inadvertent patient attrition.
;
Allocation: Randomized, Endpoint Classification: Safety/Efficacy Study, Intervention Model: Parallel Assignment, Masking: Single Blind (Outcomes Assessor), Primary Purpose: Health Services Research
| Status | Clinical Trial | Phase | |
|---|---|---|---|
| Active, not recruiting |
NCT04580030 -
Tricuapid Annular Plane Sistolic Excursion Before General Anesthesia Can Predict Hypotension After Induction
|
||
| Active, not recruiting |
NCT04279054 -
Decreased Neuraxial Morphine After Cesarean Delivery
|
Early Phase 1 | |
| Completed |
NCT03640442 -
Modified Ramped Position for Intubation of Obese Females.
|
N/A | |
| Recruiting |
NCT04099693 -
A Prospective Randomized Study of General Anesthesia Versus Anesthetist Administered Sedation for ERCP
|
||
| Terminated |
NCT02481999 -
Pre- and Postoperative EEG-Monitoring for Children Aged From 0,5 to 8 Years
|
||
| Completed |
NCT04235894 -
An Observer Rating Scale of Facial Expression Can Predict Dreaming in Propofol Anesthesia
|
||
| Recruiting |
NCT05525104 -
The Effect of DSA on Recovery of Anaesthesia in Children (Het Effect Van DSA op Het Herstel na Anesthesie Bij Kinderen).
|
N/A | |
| Recruiting |
NCT05024084 -
Desflurane and Sevoflurane Minimal Flow Anesthesia on Recovery and Anesthetic Depth
|
Phase 4 | |
| Completed |
NCT04204785 -
Noise in the OR at Induction: Patient and Anesthesiologists Perceptions
|
N/A | |
| Completed |
NCT03277872 -
NoL, HR and MABP Responses to Tracheal Intubation Performed With MAC Blade Versus Glidescope
|
N/A | |
| Terminated |
NCT03940651 -
Cardiac and Renal Biomarkers in Arthroplasty Surgery
|
Phase 4 | |
| Terminated |
NCT02529696 -
Measuring Sedation in the Intensive Care Unit Using Wireless Accelerometers
|
||
| Completed |
NCT05346588 -
THRIVE Feasibility Trial
|
Phase 3 | |
| Terminated |
NCT03704285 -
Development of pk/pd Model of Propofol in Patients With Severe Burns
|
||
| Recruiting |
NCT05259787 -
EP Intravenous Anesthesia in Hysteroscopy
|
Phase 4 | |
| Completed |
NCT02894996 -
Does the Response to a Mini-fluid Challenge of 3ml/kg in 2 Minutes Predict Fluid Responsiveness for Pediatric Patient?
|
N/A | |
| Completed |
NCT05386082 -
Anesthesia Core Quality Metrics Consensus Delphi Study
|
||
| Terminated |
NCT03567928 -
Laryngeal Mask in Upper Gastrointestinal Procedures
|
N/A | |
| Recruiting |
NCT06074471 -
Motor Sparing Supraclavicular Block
|
N/A | |
| Completed |
NCT04163848 -
CARbon Impact of aNesthesic Gas
|