Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT02815618
Other study ID # BabyLux CIP 4.1
Secondary ID
Status Completed
Phase
First received
Last updated
Start date June 2016
Est. completion date December 2022

Study information

Verified date March 2023
Source Rigshospitalet, Denmark
Contact n/a
Is FDA regulated No
Health authority
Study type Observational

Clinical Trial Summary

Feasibility trial on the use of a hybrid optical device integrating time-resolved near-infrared spectroscopy (TRS) and diffuse correlation spectroscopy (DCS) for measurement of cerebral oxygen metabolism and blood flow in neonates. The device wil be tested in four settings measuring: 1. Changes in cerebral oxygenation and haemodynamics after birth 2. precision and repeatability 3. The cerebral vaso-reactivity to arterial carbon dioxide 4. Assessment of the user-friendliness and loss of signal in routine care


Description:

The BabyLux project aims to provide a precise, accurate, and robust device to continuously monitor cerebral oxygen metabolism and blood flow in critically ill newborn infants. This will be achieved by combining time resolved near-infrared spectroscopy (TRS) with newly developed diffuse correlation spectroscopy (DCS) in a single device. The innovative aspects of the project are related to the use of advanced solutions, based on state-of-the-art photonic components, which have already been tested in laboratory and clinical tests on adults. Time Resolved Near-infrared spectroscopy and Diffuse Correlation Spectroscopy The proposed solution will integrate two advanced photonic techniques, TRS and DCS. Both techniques rely on the use of an optical fibre probe (sensor) to illuminate with very low power near-infrared light the scalp and to collect the diffusively reflected optical signal that has propagated through the scalp and skull, and therefore carries information on the deeper cortical region. The different absorption spectra of oxygenated and deoxygenated haemoglobin in the near-infrared range allows for the non-invasive monitoring of the two species in the cortical tissue. TRS and DCS prototypes are available and have been technically tested in laboratory settings and successfully validated during preclinical trials on adult volunteers and patients. Measured TRS/DCS parameters TRS measures the attenuation and the temporal broadening of relatively short light pulses (pulse duration ~100 ps) through a diffusive medium (e.g. a neonate's head). TRS has the ability to resolve path-lengths (or equivalently time-of-flights) of photons that have propagated through the tissues. This enables TRS to separate the absorption and scattering coefficients allowing for absolute measurements, and to utilize time-gating of path-lengths to emphasize signals from deeper tissues. This is particularly important for separating intra- and extra-cerebral signals for brain monitoring. DCS relies on the fact that temporal correlation of light fields in turbid media also obeys a diffusion equation, albeit a slightly different one than is used for TRS. Thus DCS shares the light penetration advantages of TRS, but, since DCS explicitly measures red blood cell movement, it provides a direct measure of quantities such as cerebral blood flow (CBF). The specific combination of DCS and TRS allows for the assessment of cerebral oxygen metabolism and CBF in a complete (i.e. CBF and oxygenation are simultaneously and independently provided), accurate (i.e. based on absolute measurements of optical parameters) and robust (i.e. potentially less affected by artefacts related to superficial systemic activity or sensor/head movements) way. The aim of this study is to perform clinical measurements using the BabyLux instrument in different clinical real-life settings to validate this new technology in terms of feasibility, repeatability of measurements, and user friendliness in neonatal medicine. The BabyLux system is tested in four different real-life settings to measure: 1. The increase in oxygenation and change in blood flow during the minutes after birth to specify the expected range of measurement; 2. The precision and repeatability of measurements by reapplying the NIRS sensor several times on slightly different sites of the head in a relatively steady condition; 3. The cerebral vaso-reactivity to arterial carbon dioxide tension in mechanically ventilated newborns to monitor induced changes in cerebral blood flow; 4. The user-friendliness and loss of signal in routine care situations (e.g. during 24-hour monitoring of neonates undergoing intensive care).


Recruitment information / eligibility

Status Completed
Enrollment 48
Est. completion date December 2022
Est. primary completion date January 2021
Accepts healthy volunteers No
Gender All
Age group N/A to 4 Weeks
Eligibility Setting 1 and 2: Inclusion Criteria: - GA > 37 weeks - planned to be delivered by an uncomplicated elective caesarean section Exclusion Criteria: - need for resuscitation or supplementary oxygen during the first 10 minutes following umbilical cord clamping - congenital malformations Setting 3: Inclusion Criteria: - GA < 37 weeks - Postnatal age more > 24 hours - Mechanically ventilated - Clinically stable - Normal brain ultrasound - Transcutaneous pCO2 monitoring (tcpCO2) Exclusion Criteria: - Congenital malformations Setting 4: Inclusion Criteria: - Postnatal age < 28 days - ventilatory support by mechanical ventilation or nasal CPAP Exclusion Criteria: - Congenital malformations

Study Design


Related Conditions & MeSH terms


Intervention

Device:
BabyLux Neuro-monitor
Measurement of cerebral blood flow index (CBFi) and tissue oxygen saturation (StO2).

Locations

Country Name City State
Denmark Rigshospitalet Copenhagen
Italy IRCCS Ca'Granda Ospedale Maggiore Policlinico Milan

Sponsors (1)

Lead Sponsor Collaborator
Gorm Greisen

Countries where clinical trial is conducted

Denmark,  Italy, 

Outcome

Type Measure Description Time frame Safety issue
Primary Cerebral tissue oxygen saturation (StO2) after birth. Measurement of cerebral haemodynamics immediately after birth. 10 min immediately after umbilical cord clamping.
Primary Precision and repeatability Test-retest variability estimated by within-subject standard deviation in one-way ANOVA with subject as factor. During second day of life.
Primary Cerebral vaso-reactivity to arterial carbon dioxide Mean CBFi and tcpCO2 one minute before the change and 15 min after will be used to analyse CBFi-tcpCO2 reactivity. 1 hour after change in ventilator settings.
Primary Assessment of user-friendliness and loss of signal in routine care Assessed by Likert-scale questionnaire completed by clinical staff. 24 hours of contineous measurements.
See also
  Status Clinical Trial Phase
Completed NCT03653910 - Effect-site Concentration of Remifentanil for Double-lumen Tubes Intubation: Airtraq VS Macintosh Laryngoscope N/A
Completed NCT06000098 - Consol Time and Acute Kidney Injury in Robotic-assisted Prostatectomy
Recruiting NCT02565251 - Volemic Resuscitation in Sepsis and Septic Shock N/A
Terminated NCT02365688 - Hemodynamic Response During Goal Directed Fluid Therapy in the OR N/A
Completed NCT01559675 - Trial Comparing Low Dose and High Dose Steroids in Patients Undergoing Colorectal Surgery N/A
Completed NCT01605279 - Dobutamine Versus Placebo for Low Superior Vena Cava Flow in Newborns Phase 2
Recruiting NCT04010058 - Postoperative Continuous Non-invasive Haemodynamic Monitoring on the Ward N/A
Completed NCT06268275 - Comparison of Effects of Scalp Block and Intravenous Esmolol on Hemodynamic Response Following the Skull Pins Application for Elective Supratentorial Craniotomy N/A
Recruiting NCT04926220 - Dynamic Estimation of Cardiac Output in the Operating Room
Not yet recruiting NCT05003011 - Correlation of Hemodynamics Via Pulmonary Artery Catherization and the Cardiospire
Completed NCT03599440 - Effect of Extended Infusion Lines on Pulse Contour-based Measurements
Completed NCT05035485 - Maternal Cardiac Output Response to Rescue Norepinephrine and Phenylephrine Boluses in Patients With Severe Preeclampsia N/A
Not yet recruiting NCT04227821 - Hemodynamic Optimalization in Pediatric Patients
Completed NCT05481047 - Prevention or Treatment of Arterial Hypotension and Oxygen Cerebral Saturation During Major Abdominal Surgery N/A
Completed NCT06158165 - Evaluating The Cardiovascular Effects of Tourniquet Application
Not yet recruiting NCT04076709 - Cardiovascular Effects of Muscle Relaxation During Laparoscopic Surgery Phase 4
Not yet recruiting NCT03828032 - Multi-parameters'Change Process During Dehydration Therapy on Brain Edema Patients. N/A
Completed NCT04089098 - VOLume and Vasopressor Therapy in Patients With Hemodynamic instAbility
Recruiting NCT05570682 - Manually Controlled Infusion vs Target Controlled Infusion for StrokeThrombectomy N/A
Completed NCT04574908 - A Pilot Trial of Continuous Portable Postoperative Hemodynamic And Saturation Monitoring On Hospital Wards N/A