Background When a patient is anesthetised and undergoes an operation, their body temperature falls. Managing a body temperature properly by keeping it within normal limits has been shown to improve recovery from surgery.

A vital part of managing temperature is being able to measure it properly. There are several ways of doing so, which consist of invasive methods such as temperature probes in the oesophagus and bladder, and non invasive methods such as infrared scanners and ear canal probes. During an operation, temperature changes as small as 0.5 °C can affect outcomes but, currently, only invasive methods are reliable enough to detect this. Oesophageal probes can't be used in patients who are not anaesthetised during their operations, and it is inappropriate to insert a urinary catheter should it not be otherwise needed.

The purpose of this study is to look at two new methods of measuring a patient's temperature which would help us in these situations. The first is called zero-flux thermometry and involves placing an adhesive pad to the forehead, which gives constant measurement of a patient's core temperature. The second involves swallowing an ingestible capsule that is no bigger than a normal pill. This technique has been very useful in research tracking a patient's core temperature as they are transferred through different areas over a longer period of time. Whilst these two methods are potentially very useful, they have never been tested on hospital patients.

Aims:

The investigators want to determine whether readings from these thermometers are accurate enough to be used in patients before, during and after surgery. Should they prove to be sufficiently accurate it will allow us to improve patient care through both clinical practice and research.

Plan:

During operations, temperature is routinely measured in line with NICE guidelines with a thermometer placed in the nose or oesophagous. The investigators will not be changing routine practice in any way, but checking measurements from the new devices against the proven, routine ones.

The investigators plan to record measurements from each device every minute throughout the operation in 20 patients. The investigators will then perform statistical tests to check their accuracy.

Study Hypothesis: Zero flux and ingestible thermometers are not significantly different at detecting patients body temperature than oesophageal temperature probes


Clinical Trial Description

Summary:

The study has been designed to provide information on the suitability and accuracy of two methods of temperature measurement in the perioperative period.

Background:

Good perioperative temperature management is an important means of improving patient outcome [1].

An essential part of optimal management is accurate temperature measurement. The gold standard is pulmonary artery catheter (PAC) measurement. However, these are used infrequently in British practice due to the high rates of associated complications. Urinary bladder and oesophageal temperatures are sufficiently accurate but are unsuitable for many patients [2]. For example it is not possible to use oesophageal probes in awake patients and many patients are not catheterised.

Although non-invasive and relatively cheap means of measuring temperature (such as aural canal probes and infrared scanners) exist, they are neither accurate nor consistent enough to distinguish the 0.5˚C differences which have a significant effect on patient outcomes [2].

In this study, the investigators plan to look at two methods of temperature measurement which potentially offer significant improvements to surgical care.

The first is a new method known as zero-flux thermometry. This is completely non-invasive, involving no more than placing an adhesive pad to the forehead. Connected up to a monitor, this will equilibrate with the patient's core temperature and give a constant read-out.

The second is an ingestible capsule. These have become an important tool for measuring both hypothermia and hyperthermia in other settings. Their accuracy has not, however, been tested in hospital patients. This technology, although too expensive for routine use, offers significant advantages for research as they can provide a constant record of temperature throughout the patient's stay in hospital. For example, the investigators' audit data show that if patients admitted with fractured neck of femur become hypothermic at any time, their mortality increases. Subsequent to this we showed that patient temperatures drop significantly between ambulance, A&E and admission to the ward. These parts of the patient journey are possible targets for the prevention of hypothermia and the continuous monitoring offered by these devices will allow us to accurately test the efficacy of the interventions.

Aims and objectives:

The aim of this study is to determine the accuracy in the perioperative period of zero-flux and ingestible temperature readings.

Previous studies have shown that oesophageal and nasopharyngeal measurements can be taken as an accurate reflection of core temperature [2]. The consistency and accuracy of the other two will be compared to this.

Plan of investigation:

This study will be purely observational. There will be no interference with current practice. Nasopharyngeal/oesophageal probes are routinely placed, according to NICE guidelines, in these patients. The zero-flux thermometer is non-invasive and the ingestible capsule no bigger than a pill, so neither provide any additional risk to the patient.

Temperature readings will be downloaded from each of the three devices with a maximum interval of 1 minute. The investigators have calculated that readings from 15 subjects will provide sufficient power. The investigators plan to recruit 20 patients. Analysis of agreement between the methods will be performed as suggested by Bland and Altman[3]. ;


Study Design

Time Perspective: Cross-Sectional


Related Conditions & MeSH terms


NCT number NCT02121574
Study type Observational
Source Brighton and Sussex University Hospitals NHS Trust
Contact
Status Completed
Phase N/A
Start date September 2014
Completion date June 2015

See also
  Status Clinical Trial Phase
Recruiting NCT05583916 - Same Day Discharge for Video-Assisted Thoracoscopic Surgery (VATS) Lung Surgery N/A
Completed NCT03213314 - HepaT1ca: Quantifying Liver Health in Surgical Candidates for Liver Malignancies N/A
Enrolling by invitation NCT05534490 - Surgery and Functionality in Older Adults N/A
Recruiting NCT04792983 - Cognition and the Immunology of Postoperative Outcomes
Recruiting NCT04612491 - Pre-operative Consultation on Patient Anxiety and First-time Mohs Micrographic Surgery
Recruiting NCT04444544 - Quality of Life and High-Risk Abdominal Cancer Surgery
Completed NCT04204785 - Noise in the OR at Induction: Patient and Anesthesiologists Perceptions N/A
Completed NCT03432429 - Real Time Tissue Characterisation Using Mass Spectrometry REI-EXCISE iKnife Study
Completed NCT04176822 - Designing Animated Movie for Preoperative Period N/A
Recruiting NCT05370404 - Prescribing vs. Recommending Over-The-Counter (PROTECT) Analgesics for Patients With Postoperative Pain: N/A
Not yet recruiting NCT05467319 - Ferric Derisomaltose/Iron Isomaltoside and Outcomes in the Recovery of Gynecologic Oncology ERAS Phase 3
Recruiting NCT04602429 - Children's Acute Surgical Abdomen Programme
Completed NCT03124901 - Accuracy of Noninvasive Pulse Oximeter Measurement of Hemoglobin for Rainbow DCI Sensor N/A
Completed NCT04595695 - The Effect of Clear Masks in Improving Patient Relationships N/A
Recruiting NCT06103136 - Maestro 1.0 Post-Market Registry
Completed NCT05346588 - THRIVE Feasibility Trial Phase 3
Completed NCT04059328 - Novel Surgical Checklists for Gynecologic Laparoscopy in Haiti
Recruiting NCT03697278 - Monitoring Postoperative Patient-controlled Analgesia (PCA) N/A
Completed NCT03355547 - Observation of Atelectasis Using Lung Ultrasonography in Children Undergoing General Anesthesia: the Cohort Study for Evaluation of the Relationship Between the Incidence and Severity of Upper Respiratory Tract Infection and the Magnitude of Anesthesia-induced Atelectasis
Recruiting NCT04776954 - Comparison of Normothermia Maintenance Between Resistive Blanket and Forced Air Warming Systems in Renal Transplant Surgery N/A